We fabricate two-layer (TL) silicon nanowires (NW) field-effect transistors (FETs) with a liquid gate. The NW devices show advanced characteristics, which reflect reliable single-electron phenomena. A strong modulation effect of channel conductivity with effectively tuned parameters is revealed. The effect opens up prospects for applications in several research fields including bioelectronics and sensing applications. Our results shed light on the nature of single trap dynamics which parameters can be fine-tuned to enhance the sensitivity of liquid-gated TL silicon nanowire FETs.
Liquid-gated Si nanowire field-effect transistor (FET) biosensors are fabricated using a complementary metal-oxide-semiconductor-compatible top-down approach. The transport and noise properties of the devices reflect the high performance of the FET structures, which allows label-free detection of cardiac troponin I (cTnI) molecules. Moreover, after removing the troponin antigens the structures demonstrate the same characteristics as before cTnI detection, indicating the reusable operation of biosensors. Our results show that the additional noise is related to the troponin molecules and has characteristics which considerably differ from those usually recorded for conventional FETs without target molecules. We describe the origin of the noise and suggest that noise spectroscopy represents a powerful tool for understanding molecular dynamic processes in nanoscale FET-based biosensors.
In the present study, transport properties and single trap phenomena in silicon nanowire (NW) field-effect transistors (FETs) are reported. The dynamic behavior of drain current in NW FETs studied before and after gamma radiation treatment deviates from the predictions of the Shockley-Read-Hall model and is explained by the concept taking into account an additional energy barrier in the accumulation regime. It is revealed that dynamics of charge exchange processes between single trap and nanowire channel strongly depend on gamma radiation treatment. The results represent potential for utilizing single trap phenomena in a number of advanced devices.
Numerous sensitive nanobiosensors are reported for various bioassay applications as a result of the development of materials science and nanotechnology. Among these sensors, nanowire (NW) field‐effect transistors (FETs) represent one of the most promising practical biosensors for ultrasensitive clinical diagnostic tools. Most studies mainly focus on how to achieve a lower detection limit but pay less attention to the long settling time effect for the detection of very small concentrations of molecules in a solution. In this study, single silicon NW FETs with long‐term stability is fabricated to investigate the settling time process at small concentrations of cardiac biomarkers relevant to myocardial diseases. It is found that the settling time strongly depends on the type of molecule, its charge state and analyte concentrations. For low concentrations, the time for measurement signals to settle down is relatively long. Therefore, it is essential to understand the settling time effect in Si NW FET‐based biosensing processes to ensure the accuracy and reliability of the detection signal. An alternative approach is demonstrated to circumvent the long measurement time by utilizing reaction kinetics parameters for the fast determination of low‐concentration detection, which also benefits the optimal balance between suitable detection time and reliable detection results.
C-reactive protein (CRP) and cardiac troponin I (cTnI) biomolecules represent the earliest enzymes that appear in the blood when a cardiac injury occurs. Real-time and selective detection of these biomarkers is essential for the prediction and detection of cardiovascular diseases at an early stage. Here we report on the label-free specific detection of both proteins at picomolar concentrations using fabricated nanowire-based biosensors. We demonstrate a novel functionalization technique based on the attachment of dibenzocyclooctyne (DBCO)-linked troponin-specific aptamers to azide-functionalized silicon (Si) nanowire (NW) surface. Due to the fast and reliable immobilization of cTnI-specific aptamers and CRP-specific antibodies on the Si NWs, the fabricated devices can rapidly detect target biomolecules demonstrating high sensitivity. We confirm the attachment of proteins to the surface of Si NWs by atomic force microscopy (AFM). Moreover, we demonstrate that nanowire structures of different sizes enable the detection of biomarkers in a wide concentration range (from 1 pg/ml to 1 µg/ml), corresponding to CRP and cTnI elevation levels during the early stage of disease formation.
Transistor biosensors are mass-fabrication-compatible devices of interest for point of care diagnosis as well as molecular interaction studies. While the actual transistor gates in processors reach the sub-10 nm range for optimum integration and power consumption, studies on design rules for the signal-to-noise ratio (S/N) optimization in transistor-based biosensors have been so far restricted to 1 µm
2
device gate area, a range where the discrete nature of the defects can be neglected. In this study, which combines experiments and theoretical analysis at both numerical and analytical levels, we extend such investigation to the nanometer range and highlight the effect of doping type as well as the noise suppression opportunities offered at this scale. In particular, we show that, when a single trap is active near the conductive channel, the noise can be suppressed even beyond the thermal limit by monitoring the trap occupancy probability in an approach analog to the stochastic resonance effect used in biological systems.
In small‐area transistors, the trapping/detrapping of charge carriers to/from a single trap located in the gate oxide near the Si/SiO2 interface leads to the discrete switching of the transistor drain current, known as single‐trap phenomena (STP), resulting in random telegraph signals. Utilizing the STP‐approach, liquid‐gated (LG) nanowire (NW) field‐effect transistor biosensors have recently been proposed for ultimate biosensing with enhanced sensitivity. In this study, the impact of channel doping concentration on the capture process of charge carriers by a single trap in LG silicon NW structures is investigated. A significant effect of the channel doping concentration on the single‐trap dynamic is revealed. To understand the mechanism behind unusual capture time behavior compared to that predicted by the classical Shockley–Read–Hall theory, an analytical model based on the rigorous description of the additional energy barrier that charge carriers have to overcome to be captured by the trap at different gate voltages is developed. The enhancement of the sensitivity for single‐trap phenomena biosensing with an increase of the channel doping concentration is explained within the framework of the proposed analytical model. The results open prospects for the development of advanced single trap‐based devices.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.