This work aims to develop analytical methods using high-performance liquid chromatography with a diode array detector (HPLC-DAD) for analysis and quantification of avermectins (AVMs) and milbemycins (MBMs) in bulk samples. First, the methods were optimized and some parameters such as temperature, flow rate, injection volume and mobile phase with different proportions of solvents were evaluated. The best chromatographic conditions were obtained using the Phenomenex ® C18 (150 × 4.60 mm, 5 μm) column at a temperature of 20 °C, flow rate of 1.2 mL min -1 , injection volume of 20 μL, and detection at 250 nm. Acetonitrile: ultrapure water (87: 13, v/v) was used as mobile phase for moxidectin and eprinomectin, and acetonitrile: methanol: ultrapure water (53: 35: 12, v/v/v) for abamectin and ivermectin. Under these conditions satisfactory results were obtained, with appropriate limits of detection and quantification, acceptable linearity, precision, accuracy, and robustness. These methods satisfy the need for analytical methods for the multi-determination of MBMs and the B1a and B1b forms of AVMs by HPLC-DAD, which can be considered simple, effective, innovative and should aid in the development of the field.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.