iFR and FFR had equivalent agreement with classification of coronary stenosis severity by HSR. Further reduction in resistance by the administration of adenosine did not improve diagnostic categorization, indicating that iFR can be used as an adenosine-free alternative to FFR.
13are common morphological findings in fatal cases of LST after DES.9-14 DES can interfere with this physiological healing process through different mechanisms: first, the antiproliferative drug released by the device prevents the cellular mitosis required to restore the endothelial continuity; second, the polymer carrying the drug exerts a proinflammatory effect itself 15 ; finally, in Background-Lack of re-endothelialization and neointimal coverage on stent struts has been put forward as the main underlying mechanism leading to late stent thrombosis. Incomplete stent apposition (ISA) has been observed frequently in patients with very late stent thrombosis after drug eluting stent implantation, suggesting a role of ISA in the pathogenesis of this adverse event. The aim of this study was to evaluate the impact of different degrees of ISA severity on abnormal shear rate and healing response with coverage, because of its potential implications for stent optimization in clinical practice. Methods and Results-We characterized flow profile and shear distribution in different cases of ISA with increasing strutwall detachment distance (ranging from 100 to 500 μm). Protruding strut and strut malapposed with moderate detachment (ISA detachment distance <100 μm) have minimal disturbance to blood flow as compared with floating strut that has more significant ISA distance. In vivo impact on strut coverage was assessed retrospectively using optical coherence tomography evaluation on 72 stents (48 patients) sequentially at baseline and after 6-month follow-up. Analysis of coverage revealed an important impact of baseline strut-wall ISA distance on the risk of incomplete strut coverage at follow-up. Malapposed segments with an ISA detachment <100 μm at baseline showed complete strut coverage at follow-up, whereas segments with a maximal ISA detachment distance of 100 to 300 μm and >300 μm had 6.1% and 15.7% of their struts still uncovered at follow-up, respectively (P<0.001). Conclusions-Flow disturbances and risk of delayed strut coverage both increase with ISA detachment distance. Insights from this study are important for understanding malapposition as a quantitative, rather than binary phenomenon (present or absent) and to define the threshold of ISA detachment that might benefit from optimization during stent implantation. (Circ Cardiovasc Interv. 2014;7:180-189.)
iFR measurements during continuous resting pressure wire pullback provide a physiological map of the entire coronary vessel. Before a PCI, the iFR pullback can predict the hemodynamic consequences of stenting specific stenoses and thereby may facilitate the intervention and stenting strategy.
Coronary plaque rupture is the most common cause of vessel thrombosis and acute coronary syndrome. The accurate early detection of plaques prone to rupture may allow prospective, preventative treatment; however, current diagnostic methods remain inadequate to detect these lesions. Established imaging features indicating vulnerability do not confer adequate specificity for symptomatic rupture. Similarly, even though experimental and computational studies have underscored the importance of endothelial shear stress in progressive atherosclerosis, the ability of shear stress to predict plaque progression remains incremental. This review examines recent advances in image-based computational modelling that have elucidated possible mechanisms of plaque progression and rupture, and potentially novel features of plaques most prone to symptomatic rupture. With further study and clinical validation, these markers and techniques may improve the specificity of future culprit plaque detection.
Stent selection based on stent model design may be critical, particularly for treatment of large artery and left main bifurcations where overexpansion is normally required to optimise results and ensure full expansion of the stent.
Background—
Coronary flow reserve has extensive validation as a prognostic marker in coronary disease. Although pressure-only fractional flow reserve (FFR) improves outcomes compared with angiography when guiding percutaneous coronary intervention, it disagrees with coronary flow reserve classification 30% of the time. We evaluated whether baseline instantaneous wave-free ratio (iFR) could provide an improved pressure-only estimation of underlying coronary flow reserve.
Methods and Results—
Invasive pressure and flow velocity were measured in 216 stenoses from 186 patients with coronary disease. The diagnostic relationship between pressure-only indices (iFR and FFR) and coronary flow velocity reserve (CFVR) was compared using correlation coefficient and the area under the receiver operating characteristic curve. iFR showed a stronger correlation with underlying CFVR (iFR–CFVR, ρ=0.68 versus FFR–CFVR, ρ=0.50;
P
<0.001). iFR also agreed more closely with CFVR in stenosis classification (iFR area under the receiver operating characteristic curve, 0.82 versus FFR area under the receiver operating characteristic curve, 0.72;
P
<0.001, for a CFVR of 2). The closer relationship between iFR and CFVR was found for different CFVR cutoffs and was particularly marked in the 0.6 to 0.9 FFR range. Hyperemic FFR flow was similar to baseline iFR flow in functionally significant lesions (FFR ≤0.75; mean FFR flow, 25.8±13.7 cm/s versus mean iFR flow, 21.5±11.7 cm/s;
P
=0.13). FFR flow was higher than iFR flow in nonsignificant stenoses (FFR >0.75; mean FFR flow, 42.3±22.8 cm/s versus mean iFR flow, 26.1±15.5 cm/s;
P
<0.001).
Conclusions—
When compared with FFR, iFR shows stronger correlation and better agreement with CFVR. These results provide physiological evidence that iFR could potentially be used as a functional index of disease severity, independently from its agreement with FFR.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.