This comprehensive core laboratory analysis comparing iFR and Pd/Pa with FFR demonstrated an overall accuracy of ~80% for both nonhyperemic indices, which can be improved to ≥90% in a subset of lesions. Clinical outcome studies are required to determine whether the use of iFR or Pd/Pa might obviate the need for hyperemia in selected patients.
iFR and FFR had equivalent agreement with classification of coronary stenosis severity by HSR. Further reduction in resistance by the administration of adenosine did not improve diagnostic categorization, indicating that iFR can be used as an adenosine-free alternative to FFR.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.