BACKGROUNDCoronary revascularization guided by fractional flow reserve (FFR) is associated with better patient outcomes after the procedure than revascularization guided by angiography alone. It is unknown whether the instantaneous wave-free ratio (iFR), an alternative measure that does not require the administration of adenosine, will offer benefits similar to those of FFR.
METHODSWe randomly assigned 2492 patients with coronary artery disease, in a 1:1 ratio, to undergo either iFR-guided or FFR-guided coronary revascularization. The primary end point was the 1-year risk of major adverse cardiac events, which were a composite of death from any cause, nonfatal myocardial infarction, or unplanned revascularization. The trial was designed to show the noninferiority of iFR to FFR, with a margin of 3.4 percentage points for the difference in risk.
RESULTSAt 1 year, the primary end point had occurred in 78 of 1148 patients (6.8%) in the iFR group and in 83 of 1182 patients (7.0%) in the FFR group (difference in risk, −0.2 percentage points; 95% confidence interval [CI], −2.3 to 1.8; P<0.001 for noninferiority; hazard ratio, 0.95; 95% CI, 0.68 to 1.33; P = 0.78). The risk of each component of the primary end point and of death from cardiovascular or noncardiovascular causes did not differ significantly between the groups. The number of patients who had adverse procedural symptoms and clinical signs was significantly lower in the iFR group than in the FFR group (39 patients [3.1%] vs. 385 patients [30.8%], P<0.001), and the median procedural time was significantly shorter (40.5 minutes vs. 45.0 minutes, P = 0.001).
CONCLUSIONSCoronary revascularization guided by iFR was noninferior to revascularization guided by FFR with respect to the risk of major adverse cardiac events at 1 year. The rate of adverse procedural signs and symptoms was lower and the procedural time was shorter with iFR than with FFR. ( Use of Instantaneous Wave-free R atio in PCI F or the past 20 years, physiological measurements obtained during invasive procedures have been used to guide coronary revascularization. Pioneering work supported the use of flow measurements to make safe decisions about revascularization, 1,2 but this approach was soon superseded by the use of fractional flow reserve (FFR), which measures pressure as a surrogate of flow to estimate the severity of stenosis. 3-5 FFR was successful largely because of its technical simplicity and because clinical trials showed that it was associated with improved clinical outcomes after percutaneous coronary intervention (PCI). 6,7 Consequently, FFR is now included in the appropriate-use criteria for coronary angiography and in the American College of Cardiology-American Heart Association-European Society of Cardiology guidelines; despite these recommendations, its adoption remains limited. [8][9][10] FFR must be measured during maximal hyperemia, which is typically induced with the administration of a potent intravenous or intracoronary vasodilator, such as adenosine. 11 Several studies have...
SummaryBackground Symptomatic relief is the primary goal of percutaneous coronary intervention (PCI) in stable angina and is commonly observed clinically. However, there is no evidence from blinded, placebocontrolled randomised trials to show its efficacy.
Intracoronary resistance is naturally constant and minimized during the wave-free period. The instantaneous wave-free ratio calculated over this period produces a drug-free index of stenosis severity comparable to FFR. (Vasodilator Free Measure of Fractional Flow Reserve [ADVISE]; NCT01118481).
This comprehensive core laboratory analysis comparing iFR and Pd/Pa with FFR demonstrated an overall accuracy of ~80% for both nonhyperemic indices, which can be improved to ≥90% in a subset of lesions. Clinical outcome studies are required to determine whether the use of iFR or Pd/Pa might obviate the need for hyperemia in selected patients.
iFR and FFR had equivalent agreement with classification of coronary stenosis severity by HSR. Further reduction in resistance by the administration of adenosine did not improve diagnostic categorization, indicating that iFR can be used as an adenosine-free alternative to FFR.
13are common morphological findings in fatal cases of LST after DES.9-14 DES can interfere with this physiological healing process through different mechanisms: first, the antiproliferative drug released by the device prevents the cellular mitosis required to restore the endothelial continuity; second, the polymer carrying the drug exerts a proinflammatory effect itself 15 ; finally, in Background-Lack of re-endothelialization and neointimal coverage on stent struts has been put forward as the main underlying mechanism leading to late stent thrombosis. Incomplete stent apposition (ISA) has been observed frequently in patients with very late stent thrombosis after drug eluting stent implantation, suggesting a role of ISA in the pathogenesis of this adverse event. The aim of this study was to evaluate the impact of different degrees of ISA severity on abnormal shear rate and healing response with coverage, because of its potential implications for stent optimization in clinical practice. Methods and Results-We characterized flow profile and shear distribution in different cases of ISA with increasing strutwall detachment distance (ranging from 100 to 500 μm). Protruding strut and strut malapposed with moderate detachment (ISA detachment distance <100 μm) have minimal disturbance to blood flow as compared with floating strut that has more significant ISA distance. In vivo impact on strut coverage was assessed retrospectively using optical coherence tomography evaluation on 72 stents (48 patients) sequentially at baseline and after 6-month follow-up. Analysis of coverage revealed an important impact of baseline strut-wall ISA distance on the risk of incomplete strut coverage at follow-up. Malapposed segments with an ISA detachment <100 μm at baseline showed complete strut coverage at follow-up, whereas segments with a maximal ISA detachment distance of 100 to 300 μm and >300 μm had 6.1% and 15.7% of their struts still uncovered at follow-up, respectively (P<0.001). Conclusions-Flow disturbances and risk of delayed strut coverage both increase with ISA detachment distance. Insights from this study are important for understanding malapposition as a quantitative, rather than binary phenomenon (present or absent) and to define the threshold of ISA detachment that might benefit from optimization during stent implantation. (Circ Cardiovasc Interv. 2014;7:180-189.)
In a population of intermediate coronary stenoses, the classification agreement between iFR and FFR is excellent and similar to that of repeated FFR measurements in the same sample. Vasodilator-independent assessment of intermediate stenosis seems applicable and may foster adoption of coronary physiology in the catheterisation laboratory.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.