A pixel-segmented ionization chamber has been designed and built by Torino University and INFN. The detector features a 24 x 24 cm2 active area divided in 1024 independent cylindrical ionization chambers and can be read out in 500 micros without introducing dead time; the digital charge quantum can be adjusted between 100 fC and 800 fC. The sensitive volume of each single ionization chamber is 0.07 cm3. The purpose of the detector is to ease the two-dimensional (2D) verifications of fields with complex shapes and large gradients. The detector was characterized in a PMMA phantom using 60Co and 6 MV x-ray photon beams. It has shown good signal linearity with respect to dose and dose rate to water. The average sensitivity of a single ionization chamber was 2.1 nC/Gy, constant within 0.5% over one month of daily measurements. Charge collection efficiency was 0.985 at the operating polarization voltage of 400 V and 3.5 Gy/min dose rate. Tissue maximum ratio and output factor have been compared with a Farmer ionization chamber and were found in good agreement. The dose profiles have been compared with the ones obtained with an ionization chamber in water phantom for the field sizes supplied by a 3D-Line dynamic multileaf collimator. These results show that this detector can be used for 2D dosimetry of x-ray photon beams, supplying a good spatial resolution and sensibly reducing the time spent in dosimetric verification of complex radiation fields.
There is a growing interest in Gafchromic films for patient dosimetry in radiotherapy and in radiology. A new model (XR-QA) with high sensitivity to low dose was tested in this study. The response of the film to different x-ray beam energies (range 28-145 kVp with various filtrations, dose range 0-100 mGy) and to visible light was investigated, together with the after exposure darkening properties. Exposed films were digitized with a commercially available, optical flatbed scanner. A single functional form for dose versus net pixel value variation has been determined for all the obtained calibration curves, with a unique fit parameter different for each of the used x-ray beams. The film response was dependent on beam energy, with higher colour variations for the beams in the range 80-140 kVp. Different sources of uncertainties in dose measurements, governed by the digitalization process, the film response uniformity and the calibration curve fit procedure, have been considered. The overall one-sigma dose measurement uncertainty depended on the beam energy and decreased with increasing absorbed dose. For doses above 10 mGy and beam energies in the range 80-140 kVp the total uncertainty was less than 5%, whereas for the 28 kVp beam the total uncertainty at 10 mGy was about 10%. The post-exposure colour variation was not negligible in the first 24 h after the exposure, with a consequent increase in the calculated dose of about 10%. Results of the analysis of the sensitivity to visible light indicated that a short exposure of this film to ambient and scanner light during the measurements will not have a significant impact on the radiation dosimetry.
Dynamic intensity-modulated radiotherapy (D-IMRT) using the slidingwindow technique is currently applied for selected treatments of head and neck cancer at Institute for Cancer Research and Treatment of Candiolo (Turin, Italy). In the present work, a PiXel-segmented ionization Chamber (PXC) has been used for the verification of 19 fields used for four different head and neck cancers. The device consists of a 32 × 32 matrix of 1024 parallel-plate ionization chambers arranged in a square of 24 × 24 cm 2 area. Each chamber has 0.4 cm diameter and 0.55 cm height; a distance of 0.75 cm separates the centre of adjacent chambers. The sensitive volume of each single ionization chamber is 0.07 cm 3 . Each of the 1024 independent ionization chambers is read out with a custom microelectronics chip.The output factors in water obtained with the PXC at a depth of 10 cm were compared to other detectors and the maximum difference was 1.9% for field sizes down to 3 × 3 cm 2 . Beam profiles for different field dimensions were measured with the PXC and two other types of ionization chambers; the maximum distance to agreement (DTA) in the 20-80% penumbra region of a 3 × 3 cm 2 field was 0.09 cm. The leaf speed of the multileaf collimator was varied between 0.07 and 2 cm s −1 and the detector response was constant to better than 0.6%. The behaviour of the PXC was measured while varying the dose rate between 0.21 and 1.21 Gy min −1 ; the mean difference was 0.50% and the maximum difference was 0.96%. Using fields obtained with an enhanced dynamic wedge and a staircase-like (step) IMRT field, the PXC has been tested for simple 1D modulated beams; comparison with film gave a maximum DTA of 0.12 cm. The PXC was then used to check four different IMRT plans for head and neck cancer treatment: cervical chordoma, parotid, ethmoid and skull base. In the comparison of the PXC versus film and PXC versus treatment planning system, the number of pixels with γ parameter 1 was 97.7% and 97.6%, respectively.
Gafchromic XR-QA films were developed for patient dosimetry in diagnostic radiology. A possible application of these films is the measurement of doses in computed tomography. In this study a method to evaluate the CTDI using Gafchromic XR-QA film and a flatbed scanner was developed and tested. Film samples were cut to dimensions of 6 x 170 mm2 in order to have an integration area similar to that of a pencil ionization chamber, with the possibility of changing the integration length. Prior to exposing these films to a computed tomography beam, the angular dependence of the film dose response was investigated by exposing film strips to a static x-ray beam at different angles in the range 0 degrees-180 degrees. A difference of 49% was found between the response with the axis beam parallel to the film surface (90 degrees) and with the axis beam perpendicular (0 degrees and 180 degrees). Integrating over a 360 degrees exposure like the one in computed tomography, a difference of less than 2% was estimated, which is comparable with the measurement error obtainable with XR-QA film. A calibration with a CT beam in the scout mode was performed and film strips were then exposed to single axial scans and to helical scans both in air and in phantoms. Two different types of flatbed scanners were used to read the film samples, a Microtek ScanMaker 9800XL scanner and an Epson Expression 10000 XL scanner, and the accuracy of the results were compared. For beam collimations above 10 mm differences between CTDI measured by film and CTDI measured by ionization chamber below 9% were found for the Epson scanner, with an average estimated error at 1 sigma level of 5%. For the Microtek scanner and for the same film samples, differences below 11% with an average error at 1 sigma level of 8% were founded. The 1 sigma uncertainty of the measured CTDI was provided by the method for each measurement, and it was shown that about the 95% of the differences between the CTDI measurements with radiochromic films and with the ionization chamber were below the estimated 2 sigma uncertainty, for both scanners. After an accurate calibration procedure and the consideration of the uncertainty associated with the measurement, Gafchromic XR-QA films can be used to evaluate the CTDI.
The purpose of this study was to evaluate the consequences of different choices of acquisition parameters on the actual image noise and on the patient dose with an automatic tube current modulation system. The CT investigated was a GE Lightspeed 16-slice and an anthropomorphic phantom was used to simulate the patient. Several acquisitions were made varying noise index (NI), kilovoltage and pitch values. Tube current values were compared for the different acquisitions. Patient dose was evaluated in terms of volumetric computed tomography dose index (CTDI(vol)) and also as effective dose. The noise actually present in the images was analyzed by a region of interest analysis considering representatively phantom sections in the regions of the shoulders, of the lungs and of the abdomen. The obtained results generally evidenced a good agreement between the noise index and the measured noise for the abdomen sections, whereas for the shoulders and the lungs sections the measured noise was respectively greater and lower of the NI. Varying the kV the automatic current modulation system provided images with a substantially constancy of the actual noise and of the patient dose. An increase of the pitch generally decreased the patient dose, whereas the noise was slightly greater for the lowest pitch and almost constant for the other pitch values. This study outlines some important relationships between an automatic tube current modulation system and other CT acquisition parameters, providing useful informations for the choice requested by radiologists in the task of optimization of the CT acquisition protocols. Unless there are other considerations in place, pixel pitches below 1.375 should be avoided, and kVp settings can be changed with no real impact on dose or image noise.
Two detectors for fast two-dimensional (2D) and quasi-three-dimensional (quasi-3D) verification of the dose delivered by radiotherapy beams have been developed at University and Istituto Nazionale di Fisica Nucleare (INFN) of Torino. The Magic Cube is a stack of strip-segmented ionization chambers interleaved with water-equivalent slabs. The parallel plate ionization chambers have a sensitive area of 24 x 24 cm2, and consist of 0.375 cm wide and 24 cm long strips. There are a total of 64 strips per chamber. The Magic Cube has been tested with the clinical proton beam at Loma Linda University Medical Centre (LLUMC), and was shown to be capable of fast and precise quasi-3D dose verification. The Pixel Ionization Chamber (PXC) is a detector with pixel anode segmentation. It is a 32 x 32 matrix of 1024 cylindrical ionization cells arranged in a square 24 x 24 cm2 area. Each cell has 0.4 cm diameter and 0.55 cm height, at a pitch of 0.75 cm separates the centre of adjacent cells. The sensitive volume of each single ionization cell is 0.07 cm3. The detectors are read out using custom designed front-end microelectronics and a personal computer-based data acquisition system. The PXC has been used to verify dynamic intensity-modulated radiotherapy for head-and-neck and breast cancers.
We present a quick and easy method for the calibration of a matrix of sensors. The algorithm is based on a three-step irradiation procedure which relies only on the constancy of the delivered fluence at each step. With this method the gain of each sensor is derived relative to a reference detector. The algorithm has been applied to a matrix of (32 x 32) ionization chambers. After the calibration coefficients have been applied, by comparing the response of the matrix of chambers to a reference detector over a large field, we determined that the fluence measurement of individual chambers is better than 0.7%. The algorithm solves the cumbersome problem of the relative gain calibration of a matrix of a large number of sensors.
Gafchromic XR-R films are a useful tool to evaluate entrance skin dose in interventional radiology. Another dosimetric quantity of interest in diagnostic and interventional radiology is the dose area product (DAP). In this study, a method to evaluate DAP using Gafchromic XR-R films and a flat-bed scanner was developed and tested. Film samples were exposed to an x-ray beam of 80 kVp over a dose range of 0-10 Gy. DAP measurements with films were obtained from the digitalization of a film sample positioned over the x-ray beam window during the exposure. DAP values obtained with this method were compared for 23 cardiological interventional procedures with DAP values displayed by the equipment. The overall one-sigma dose measurement uncertainty depended on the absorbed dose, with values below 6% for doses above 1 Gy. A maximum discrepancy of 16% was found, which is of the order of the differences in the DAP measurements that may occur with different calibration procedures. Based on the results presented, after an accurate calibration procedure and a thorough inspection of the relationship between the actual dose and the direct measured quantity (net optical density or net pixel value variation), Gafchromic XR-R films can be used to assess the DAP.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
hi@scite.ai
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.