There is a growing interest in Gafchromic films for patient dosimetry in radiotherapy and in radiology. A new model (XR-QA) with high sensitivity to low dose was tested in this study. The response of the film to different x-ray beam energies (range 28-145 kVp with various filtrations, dose range 0-100 mGy) and to visible light was investigated, together with the after exposure darkening properties. Exposed films were digitized with a commercially available, optical flatbed scanner. A single functional form for dose versus net pixel value variation has been determined for all the obtained calibration curves, with a unique fit parameter different for each of the used x-ray beams. The film response was dependent on beam energy, with higher colour variations for the beams in the range 80-140 kVp. Different sources of uncertainties in dose measurements, governed by the digitalization process, the film response uniformity and the calibration curve fit procedure, have been considered. The overall one-sigma dose measurement uncertainty depended on the beam energy and decreased with increasing absorbed dose. For doses above 10 mGy and beam energies in the range 80-140 kVp the total uncertainty was less than 5%, whereas for the 28 kVp beam the total uncertainty at 10 mGy was about 10%. The post-exposure colour variation was not negligible in the first 24 h after the exposure, with a consequent increase in the calculated dose of about 10%. Results of the analysis of the sensitivity to visible light indicated that a short exposure of this film to ambient and scanner light during the measurements will not have a significant impact on the radiation dosimetry.
Gafchromic XR-QA films were developed for patient dosimetry in diagnostic radiology. A possible application of these films is the measurement of doses in computed tomography. In this study a method to evaluate the CTDI using Gafchromic XR-QA film and a flatbed scanner was developed and tested. Film samples were cut to dimensions of 6 x 170 mm2 in order to have an integration area similar to that of a pencil ionization chamber, with the possibility of changing the integration length. Prior to exposing these films to a computed tomography beam, the angular dependence of the film dose response was investigated by exposing film strips to a static x-ray beam at different angles in the range 0 degrees-180 degrees. A difference of 49% was found between the response with the axis beam parallel to the film surface (90 degrees) and with the axis beam perpendicular (0 degrees and 180 degrees). Integrating over a 360 degrees exposure like the one in computed tomography, a difference of less than 2% was estimated, which is comparable with the measurement error obtainable with XR-QA film. A calibration with a CT beam in the scout mode was performed and film strips were then exposed to single axial scans and to helical scans both in air and in phantoms. Two different types of flatbed scanners were used to read the film samples, a Microtek ScanMaker 9800XL scanner and an Epson Expression 10000 XL scanner, and the accuracy of the results were compared. For beam collimations above 10 mm differences between CTDI measured by film and CTDI measured by ionization chamber below 9% were found for the Epson scanner, with an average estimated error at 1 sigma level of 5%. For the Microtek scanner and for the same film samples, differences below 11% with an average error at 1 sigma level of 8% were founded. The 1 sigma uncertainty of the measured CTDI was provided by the method for each measurement, and it was shown that about the 95% of the differences between the CTDI measurements with radiochromic films and with the ionization chamber were below the estimated 2 sigma uncertainty, for both scanners. After an accurate calibration procedure and the consideration of the uncertainty associated with the measurement, Gafchromic XR-QA films can be used to evaluate the CTDI.
Chromium-51 ethylene diamine tetra-acetic acid (51Cr-EDTA) total plasma clearance was evaluated using a multi-sample method (i.e. 12 blood samples) as the reference compared with several simplified methods which necessitated only one or few blood samples. The following 5 methods were evaluated: terminal slope-intercept method with 3 blood samples, simplified method of Bröchner-Mortensen and 3 single-sample methods (Constable, Christensen and Groth, Tauxe). Linear regression analysis was performed. Standard error of estimate, bias and imprecision of different methods were evaluated. For 51Cr-EDTA total plasma clearance greater than 30 ml.min-1, the results which most approximated the reference source were obtained by the Christensen and Groth method at a sampling time of 300 min (inaccuracy of 4.9%). For clearances between 10 and 30 ml.min-1, single-sample methods failed to give reliable results. Terminal slope-intercept and Bröchner-Mortensen methods were better, with inaccuracies of 17.7% and 16.9%, respectively. Although sampling times at 180, 240 and 300 min are time-consuming for patients, 51Cr-EDTA total plasma clearance can be accurately calculated for values greater than 10 ml.min-1 using the Bröchner-Mortensen method. In patients with clearance greater than 30 ml.min-1, single-sample techniques provide a good alternative to the multi-sample method; the choice of the method to be used depends on the degree of accuracy required.
The purpose of this study was to evaluate the consequences of different choices of acquisition parameters on the actual image noise and on the patient dose with an automatic tube current modulation system. The CT investigated was a GE Lightspeed 16-slice and an anthropomorphic phantom was used to simulate the patient. Several acquisitions were made varying noise index (NI), kilovoltage and pitch values. Tube current values were compared for the different acquisitions. Patient dose was evaluated in terms of volumetric computed tomography dose index (CTDI(vol)) and also as effective dose. The noise actually present in the images was analyzed by a region of interest analysis considering representatively phantom sections in the regions of the shoulders, of the lungs and of the abdomen. The obtained results generally evidenced a good agreement between the noise index and the measured noise for the abdomen sections, whereas for the shoulders and the lungs sections the measured noise was respectively greater and lower of the NI. Varying the kV the automatic current modulation system provided images with a substantially constancy of the actual noise and of the patient dose. An increase of the pitch generally decreased the patient dose, whereas the noise was slightly greater for the lowest pitch and almost constant for the other pitch values. This study outlines some important relationships between an automatic tube current modulation system and other CT acquisition parameters, providing useful informations for the choice requested by radiologists in the task of optimization of the CT acquisition protocols. Unless there are other considerations in place, pixel pitches below 1.375 should be avoided, and kVp settings can be changed with no real impact on dose or image noise.
Quality controls for testing the performance of computed radiography (CR) systems have been recommended by manufacturers and medical physicists' organizations. The purpose of this work was to develop a set of image processing tools for quantitative assessment of computed radiography quality control parameters. Automatic image analysis consisted in detecting phantom details, defining regions of interest and acquiring measurements. The tested performance characteristics included dark noise, uniformity, exposure calibration, linearity, low-contrast and spatial resolution, spatial accuracy, laser beam function and erasure thoroughness. CR devices from two major manufacturers were evaluated. We investigated several approaches to quantify the detector response uniformity. We developed methods to characterize the spatial accuracy and resolution properties across the entire image area, based on the Fourier analysis of the image of a fine wire mesh. The implemented methods were sensitive to local blurring and allowed us to detect a local distortion of 4% or greater in any part of an imaging plate. The obtained results showed that the developed image processing tools allow us to implement a quality control program for CR with short processing time and with absence of subjectivity in the evaluation of the parameters.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.