Stereotactic radiosurgery of the brain may be accomplished with a linear accelerator by performing several noncoplanar arcs of a highly collimated beam focused at a point. The shape of the radiation distribution produced by this technique is affected by the beam energy, field size, and the number and size of the arcs. The influence of these parameters on the resulting radiation distributions was analyzed by computing dose volume histograms for a typical brain. Dose volume functions were computed for: (a) the energy range of 4-24 MV x rays; (b) target sizes of 1-4 cm; and (c) 1-11 arcs and dynamic rotation. The dose volume histograms were found to be dependent on the number of arcs for target sizes of 1-4 cm. However, these differences were minimal for techniques with 4 arcs or more. The influence of beam energy on the dose volume histogram was also found to be minimal.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.