Initial performance evaluation of the UniCel® DxH 800 Coulter® cellular analysis system

Abstract: The Beckman Coulter UniCel® DxH 800 is a hematology analyzer incorporating new electronic and mechanical design with advanced algorithm technology to perform CBC, white blood cell (WBC) differential, nucleated red blood cell (NRBC), and reticulocyte analysis. Evaluation of this instrument was performed in our 800-bed tertiary care hospital and specifically centered upon the correlation of WBC, NRBC, and platelet (PLT) enumeration when compared to a predicate analyzer, the Coulter® LH 780, and flow cytometry (F… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

2
29
0
1

Year Published

2011
2011
2021
2021

Publication Types

Select...
7
2

Relationship

0
9

Authors

Journals

citations
Cited by 30 publications
(32 citation statements)
references
References 22 publications
2
29
0
1
Order By: Relevance
“…With the exception of MPV, the r values were greater than 0.97 and comparable to those in a multi-instrument study by Muller et al 6 and a recent evaluation of the DxH 800 vs the LH 780. 1 In addition to the r value, which indicates how well the data conform to the best-fit line, additional statistical parameters include the slope and y-intercept, which indicate the degree of proportional and absolute bias, respectively. All parameters for the CBC exhibited slopes close to 1 and y-intercepts close to 0, with the exceptions of the MPV and platelet count, for which the DxH 800 values were similar to those of the LH 780 but proportionately less than those of the Sapphire (slope = 0.820 for platelets; slope = 0.850 for MPV).…”
Section: Resultsmentioning
confidence: 99%
See 1 more Smart Citation
“…With the exception of MPV, the r values were greater than 0.97 and comparable to those in a multi-instrument study by Muller et al 6 and a recent evaluation of the DxH 800 vs the LH 780. 1 In addition to the r value, which indicates how well the data conform to the best-fit line, additional statistical parameters include the slope and y-intercept, which indicate the degree of proportional and absolute bias, respectively. All parameters for the CBC exhibited slopes close to 1 and y-intercepts close to 0, with the exceptions of the MPV and platelet count, for which the DxH 800 values were similar to those of the LH 780 but proportionately less than those of the Sapphire (slope = 0.820 for platelets; slope = 0.850 for MPV).…”
Section: Resultsmentioning
confidence: 99%
“…This study represents one of the few early reports on the DxH 800, which is a recently released, high-volume hematology analyzer intended to supplant the previous generation, Beckman Coulter LH 780. [1][2][3] The DxH 800 represents a new design from the manufacturer with less tubing and a smaller footprint than the LH 780. The Cell-Dyn Sapphire represents a comparable high-volume analyzer from Abbott Diagnostics that has been studied extensively.…”
mentioning
confidence: 99%
“…The Unicel DxH 800 did not reproduce the contamination phenomenon in our hands and has superior sensitivity for NRBCs compared with the LH 750. 4 Our blood smears with unusual NRBCs did not contain other abnormalities that typically accompany NRBCs, such as immature cells or dysplastic granulocytes. Unexpected laboratory findings must be carefully evaluated in their clinical setting with careful consideration of preanalytical as well as analytical variables.…”
Section: Discussionmentioning
confidence: 86%
“…The PLT count on this analyzer has been shown to be linear in the range comprised between 1 and 1297 Â10 9 /l [5]. The blood was collected by an experience phlebotomist, using a 21 gauge straight needle, directly into evacuated blood tubes containing 109 mmol/l buffered sodium citrate (Venosafe; Terumo Europe N.V., Leuven, Belgium; ref.…”
Section: Methodsmentioning
confidence: 99%