Comparison of Visian toric collamer lens and toric acrylic intraocular lens implantation for the treatment of myopia with astigmatism

Abstract: PurposeTo compare the efficacy and outcome of phakic toric implantable collamer lens (TICL) and refractive clear lens extraction with AcrySof Toric intraocular lens (TIOL) implantation for the treatment of myopic astigmatism.Patients and methodsThis study assessed eyes with myopic astigmatism >−1 D and ≤−4 D with a spherical equivalent >10 D or <10 D if the patients were unsuitable for corneal refractive surgery. These eyes were divided into group A, in which Visian Toric ICL™ Phakic TICL was implanted, and gr… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1

Citation Types

0
3
0

Year Published

2019
2019
2021
2021

Publication Types

Select...
3

Relationship

0
3

Authors

Journals

citations
Cited by 3 publications
(3 citation statements)
references
References 47 publications
(34 reference statements)
0
3
0
Order By: Relevance
“…All patients had satisfactory improvements in their visual acuities. Three of the patients who underwent toric intra-ocular lens implantation had lens rotation few days after surgery that resulted in decrease in their visual acuity and that was overcome by realignment of their IOL intraoperatively [6,7]. They were followed for up to three months with no change in their BCVA or their lens status.…”
Section: Resultsmentioning
confidence: 99%
See 2 more Smart Citations
“…All patients had satisfactory improvements in their visual acuities. Three of the patients who underwent toric intra-ocular lens implantation had lens rotation few days after surgery that resulted in decrease in their visual acuity and that was overcome by realignment of their IOL intraoperatively [6,7]. They were followed for up to three months with no change in their BCVA or their lens status.…”
Section: Resultsmentioning
confidence: 99%
“…Three of the patients who received Visian ICLs had lens rotation after surgery and this was due to under sizing of the lens [4,6]. They underwent lens exchange and replaced with a larger ICL with more stable alignment.…”