2010
DOI: 10.1002/ccd.22615
|View full text |Cite
|
Sign up to set email alerts
|

A randomized comparison of TR band and radistop hemostatic compression devices after transradial coronary intervention

Abstract: We have shown in a randomized comparison of Radistop and TR band that both devices are safe and effective as hemostatic compression devices following transradial procedures. However, more patients felt discomfort with the Radistop device and the time taken to achieve hemostasis was longer with TR band. © 2010 Wiley-Liss, Inc.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
42
1
5

Year Published

2010
2010
2023
2023

Publication Types

Select...
6
3

Relationship

0
9

Authors

Journals

citations
Cited by 60 publications
(48 citation statements)
references
References 18 publications
0
42
1
5
Order By: Relevance
“…On the other hand, individuals allocated to the Radistop™ group required a shorter duration of arterial compression (289 min with Radistop™ and 319 min with TR Band™). The incidence of RAO was similar in the two groups (9.6% with TR Band™ vs. 8.9% with Radistop™, p= 0.9) (11) . The radial approach has been adopted worldwide as the preferred route for most interventional cardiology procedures.…”
Section: Introductionmentioning
confidence: 79%
“…On the other hand, individuals allocated to the Radistop™ group required a shorter duration of arterial compression (289 min with Radistop™ and 319 min with TR Band™). The incidence of RAO was similar in the two groups (9.6% with TR Band™ vs. 8.9% with Radistop™, p= 0.9) (11) . The radial approach has been adopted worldwide as the preferred route for most interventional cardiology procedures.…”
Section: Introductionmentioning
confidence: 79%
“…Em análise comparativa randomizada envolvendo 790 pacientes, esse dispositivo promoveu elevado porcentual de conforto ao paciente, baixas taxas de complicações vasculares e oclusão arterial, a despeito de maior tempo para obtenção da hemostasia. 14 De fato, a obediência às instruções do fabricante para a redução da pressão de compressão requer a análise inicial uma hora após o término do procedimento, seguida de avaliações a cada 30 minutos. Assim, em centros com alto volume de procedimentos, isso representaria um fator limitante ao uso irrestrito do dispositivo.…”
Section: Discussionunclassified
“…Auf dem Markt verfügbar sind unterschiedliche Verschlusssysteme. Es gibt nur wenige wissenschaftliche Arbeiten über den Vergleich der Systeme untereinander [11,38], sodass eine adä-quate Wertung hinsichtlich der Qualität nicht möglich ist.…”
Section: Nachsorgeunclassified