Coronavirus disease 2019 (COVID-19) is a highly contagious disease caused by severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2). Diagnosis of COVID-19 depends on quantitative reverse transcription PCR (qRT-PCR), which is time-consuming and requires expensive instrumentation. Here, we report an ultrasensitive electrochemical biosensor based on isothermal rolling circle amplification (RCA) for rapid detection of SARS-CoV-2. The assay involves the hybridization of the RCA amplicons with probes that were functionalized with redox active labels that are detectable by an electrochemical biosensor. The one-step sandwich hybridization assay could detect as low as 1 copy/μL of N and S genes, in less than 2 h. Sensor evaluation with 106 clinical samples, including 41 SARS-CoV-2 positive and 9 samples positive for other respiratory viruses, gave a 100% concordance result with qRT-PCR, with complete correlation between the biosensor current signals and quantitation cycle (Cq) values. In summary, this biosensor could be used as an on-site, real-time diagnostic test for COVID-19.
COVID-19 is a highly contagious disease caused by severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2). Diagnosis of COVID-19 depends on quantitative reverse transcription PCR (qRT-PCR), which is time-consuming and requires expensive instrumentation. Here, we report an ultrasensitive electrochemical biosensor based on isothermal rolling circle amplification (RCA) for rapid detection of SARS-CoV-2. The assay involves the hybridization of the RCA amplicons with probes that were functionalized with redox active labels that are detectable by an electrochemical biosensor. The one-step sandwich hybridization assay could detect as low as 1 copy/mL of N and S genes, in less than 2 hours. Sensor evaluation with 105 clinical samples, including 40 SARS-CoV-2 positive and 9 samples positive for other respiratory viruses, gave a 100% concordance result with qRT-PCR, with complete correlation between the biosensor current signals and quantitation cycle (Cq) values. In summary, this biosensor could be used as an on-site, real-time diagnostic test for COVID-19.
Infection with high-risk human papillomavirus (HPV) is a major risk factor for oral and cervical cancers. Hence, we developed a multianalyte electrochemical DNA biosensor that could be used for both oral and cervical samples to detect the high-risk HPV genotypes 16 and 18. The assay involves the sandwich hybridization of the HPV target to the silica-redox dye reporter probe and capture probe, followed by electrochemical detection. The sensor was found to be highly specific and sensitive, with a detection limit of 22 fM for HPV-16 and 20 fM for HPV-18, between the range of 1 fM and 1 µM. Evaluation with oral and cervical samples showed that the biosensor result was consistent with the nested PCR/gel electrophoresis detection. The biosensor assay could be completed within 90 min. Due to its simplicity, rapidity, and high sensitivity, this biosensor could be used as an alternative method for HPV detection in clinical laboratories as well as for epidemiological studies.
The detection and identification of multiple components in a complex sample such as food in a cost-effective way is an ongoing challenge. The development of on-site and rapid detection methods to ensure food quality and composition is of significant interest to the food industry. Here we report that an electrochemical method can be used with an unmodified glassy carbon electrode for the identification of the key ingredients found within Thai green curries. It was found that green curry presents a fingerprint electrochemical response that contains four distinct peaks when differential pulse voltammetry is performed. The reproducibility of the sensor is excellent as no surface modification is required and therefore storage is not an issue. By employing particle swarm optimization algorithms the identification of ingredients within a green curry could be obtained. In addition, the quality and freshness of the sample could be monitored by detecting a change in the intensity of the peaks in the fingerprint response.
We report a highly sensitive and selective multiplex assay by empowering an electrochemical DNA sensor with isothermal rolling circle amplification. The assay could simultaneously detect and discriminate three common entero-pathogens...
Background
Wolbachia is an endosymbiont bacterium generally found in about 40% of insects, including mosquitoes, but it is absent in Aedes aegypti which is an important vector of several arboviral diseasesdengue, chikungunya, zika, and yellow fever. The evidence that Wolbachia trans-infected Ae. aegypti mosquitoes lost their vectorial competence and became less capable of transmitting arboviruses to human hosts highlights the potential of using Wolbachia-based approaches for prevention and control of arboviral diseases. Recently, release of Wolbachia trans-infected Ae. aegypti has been deployed widely in many countries for the control of mosquito-borne viral diseases. Field surveillance and monitoring of Wolbachia presence in released mosquitoes is important for the success of these control programs. So far, a number of studies have reported the development of loop mediated isothermal amplification (LAMP) assays to detect Wolbachia in mosquitoes, but the methods still have some specificity and cost issues.
Methodology/Principal findings
We describe here the development of a LAMP assay combined with the DNA strand displacement-based electrochemical sensor (BIOSENSOR) method to detect wAlbB Wolbachia in trans-infected Ae. aegypti. Our developed LAMP primers used a low-cost dye detecting system and 4 oligo nucleotide primers which can reduce the cost of analysis while the specificity is comparable to the previous methods. The detection capacity of our LAMP technique was 1.4 nM and the detection limit reduced to 2.2 fM when combined with the BIOSENSOR. Our study demonstrates that a BIOSENSOR can also be applied as a stand-alone method for detecting Wolbachia; and it showed high sensitivity when used with the crude DNA extracts of macerated mosquito samples without DNA purification.
Conclusions/Significance
Our results suggest that both LAMP and BIOSENSOR, either used in combination or stand-alone, are robust and sensitive. The methods have good potential for routine detection of Wolbachia in mosquitoes during field surveillance and monitoring of Wolbachia-based release programs, especially in countries with limited resources.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.