In order to give the correct dose to a patient, the monitor chamber for a proton scanning system has to be calibrated. As recombination of ion pairs occurs in the monitor chamber, the relation between the number of particles traversing it per time unit and the ionization chamber signal is not linear. A method developed for a scanned pulsed proton beam taking the nonlinear monitor signal into account is described. A vital part of the reference dosimetry procedure is to determine the absorbed dose under reference conditions, which is recommended to be done with an ionization chamber. For a scanned pulsed proton beam, the recombination in the ionization chamber is not negligible and the signal from the ionization chamber has to be corrected. In this work, it is shown that although the pulse length is comparable to the ion transit time the beam can be considered as continuously scanned if the applied high voltage is not too small. Also shown is that the two-voltage formula for a continuous beam is under some conditions applicable for a continuous scanned beam as well.
The implementation of two algorithms for calculating dose distributions for radiation therapy treatment planning of intermediate energy proton beams is described. A pencil kernel algorithm and a depth penetration algorithm have been incorporated into a commercial three dimensional treatment planning system (Helax-TMS, Helax AB, Sweden) to allow conformal planning techniques using irregularly shaped fields, proton range modulation, range modification and dose calculation for non-coplanar beams. The pencil kernel algorithm is developed from the Fermi Eyges formalism and Molière multiple-scattering theory with range straggling corrections applied. The depth penetration algorithm is based on the energy loss in the continuous slowing down approximation with simple correction factors applied to the beam penumbra region and has been implemented for fast, interactive treatment planning. Modelling of the effects of air gaps and range modifying device thickness and position are implicit to both algorithms. Measured and calculated dose values are compared for a therapeutic proton beam in both homogeneous and heterogeneous phantoms of varying complexity. Both algorithms model the beam penumbra as a function of depth in a homogeneous phantom with acceptable accuracy. Results show that the pencil kernel algorithm is required for modelling the dose perturbation effects from scattering in heterogeneous media.
A scanned proton beam yields dose distributions that in most cases are superior to passively scattered proton beams and to other external radiation treatment modalities. The present paper gives a description of the scanning system that has been developed at the Svedberg Laboratory (TSL) in Uppsala. The scanning technique and the technical design are described. The solution with a small pole gap of the magnets and a moveable second magnet results in a very compact scanning head, which can therefore be incorporated in a gantry of relatively limited size. A prototype was constructed that has been used to realize various dose distributions with a scanned beam of 180 MeV protons at TSL.
In this paper we present the recent developments made for the scanning system for proton beams at TSL in Uppsala, showing that this system is now fully functional being able to produce conformal intensity modulated scan patterns with sufficient accuracy. A new control and supervising system handling the beam delivery including the control of the synchrocyclotron and the scanning system is developed and described in detail. A complete dosimetry system with transmission ionization chambers and a multi-wire ionization chamber for monitoring of the beam during scanning has been constructed. The details of the dose monitors and the position sensitive multi-wire ionization chamber are presented in this work. Furthermore, we have established procedures for verification measurements to ensure the quality of the beam and also methods for calibration of the beam monitors and relative and absolute dosimetry for complex scanned beams.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
hi@scite.ai
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.