Background: Before treatment planning and dose delivery, quality assurance of multi-leaf collimator (MLC) has an important role in intensity-modulated radiation therapy (IMRT) due to the creation of multiple segments from optimization process.Objective: The purpose of this study is to assess the quality control of MLC leaves using EBT3 Gafchromic films.Material and Methods: Leaf Position accuracy and leaf gap reproducibility were checked with Garden fence test. The garden fence test consists of 5 thin bands A) 0.2 Cm width spaced at 2 Cm intervals and B) 1 Cm width spaced at 1 Cm intervals. Each leaf accuracy was analyzed with measuring the full-width half-maximum (FWHM). Maximum and average leaf transmission were measured with gafchromic EBT3 films from Ashland for both 6 MV and 18 MV beams.Results: Leaf positions were found to be in a range between 1.78 – 2.53 mm, instead of nominal 2 mm for the test A and between 9.09 – 10.36 mm, instead of nominal 10 mm for the test B. The Average radiation transmission of the MLC was noted 1.79% and 1.98% of the open 10x10 Cm2 field at isocenter for 6 MV and 18 MV beams, respectively. Maximum radiation transmission was noted 4.1% and 4.4% for 6 MV and 18 MV beams, respectively.Conclusion: In this study, application of gafchromic EBT3 films for the quality assurance of Euromechanics multileaf collimator was studied. Our results showed that the average leaf leakage and positional accuracy of this type of MLC were in the acceptance level based on the Protocols.
Background: Before treatment planning and dose delivery, quality assurance of multi-leaf collimator (MLC) has an important role in intensity-modulated radiation therapy (IMRT) due to the creation of multiple segments from optimization process.Objective: The purpose of this study is to assess the quality control of MLC leaves using EBT3 Gafchromic films.Material and Methods: Leaf Position accuracy and leaf gap reproducibility were checked with Garden fence test. The garden fence test consists of 5 thin bands A) 0.2 Cm width spaced at 2 Cm intervals and B) 1 Cm width spaced at 1 Cm intervals. Each leaf accuracy was analyzed with measuring the full-width half-maximum (FWHM). Maximum and average leaf transmission were measured with gafchromic EBT3 films from Ashland for both 6 MV and 18 MV beams.Results: Leaf positions were found to be in a range between 1.78 – 2.53 mm, instead of nominal 2 mm for the test A and between 9.09 – 10.36 mm, instead of nominal 10 mm for the test B. The Average radiation transmission of the MLC was noted 1.79% and 1.98% of the open 10x10 Cm2 field at isocenter for 6 MV and 18 MV beams, respectively. Maximum radiation transmission was noted 4.1% and 4.4% for 6 MV and 18 MV beams, respectively.Conclusion: In this study, application of gafchromic EBT3 films for the quality assurance of Euromechanics multileaf collimator was studied. Our results showed that the average leaf leakage and positional accuracy of this type of MLC were in the acceptance level based on the Protocols.
In this study, the beam characteristics of a Euromechanics add-on MLC that has been installed on a Varian CLINAC 2100 C/D linear accelerator are presented. This was the first installation of 60-leaf PMLC from Euromechanics Company worldwide and all mechanical and dosimetric parameters were measured before clinical use of this kind of MLC. Mechanical tests were executed for different gantry and collimator angles. Leaf position accuracy and leaf gap reproducibility were checked with four different tests. The leaf transmissions, collimator (Sc), phantom (Sp), total (Sc,p) scatter factors, output of the machine, beam profiles for off-axis ratios, central axis depth dose, flatness, symmetry and penumbra have been measured for different field sizes pre and post MLC installation in 6 and 18 MV-mode. To evaluate the effect of new data on clinical plans, different beam setup configurations conformed with MLC and custom blocks were planned on CT images of thorax a CIRS phantom model 002LFC in the same treatment planning system. Leaf position in picket fence test found to be in range between 4.89–5.02 cm instead of nominal 5 cm, however the results of this test with EPIDs image and PIPSpro software showed the higher deviation rather than the results reported from the tests with EBT3 films. The measured data showed that on average Sc,p and Sc were increased 0.22% (P = 0.86) and 0.34% (P = 0.86) for 6 MV and 0.37% (P = 0.84) and 0.42% (P = 0.88) for 18 MV beams for different field sizes, respectively. Good agreement was observed between the PDD and profile curves pre and post MLC installation that was expected based on no changes in beam energy and geometry of the collimators. Based on the mechanical and dosimetry results which have been achieved from our different standard tests, it was found no significant differences between pre and post MLC installation values. This indicates, installation and using this system is clinically acceptable.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
hi@scite.ai
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.