2019
DOI: 10.31661/jbpe.v0i0.1045
|View full text |Cite
|
Sign up to set email alerts
|

Physical and Dosimetric Aspect of Euromechanics Add-on Multileaf Collimator on Varian Clinac 2100 C/D

Abstract: Background: Before treatment planning and dose delivery, quality assurance of multi-leaf collimator (MLC) has an important role in intensity-modulated radiation therapy (IMRT) due to the creation of multiple segments from optimization process.Objective: The purpose of this study is to assess the quality control of MLC leaves using EBT3 Gafchromic films.Material and Methods: Leaf Position accuracy and leaf gap reproducibility were checked with Garden fence test. The garden fence test consists of 5 thin bands A)… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
2

Citation Types

0
10
0

Year Published

2023
2023
2024
2024

Publication Types

Select...
1
1

Relationship

0
2

Authors

Journals

citations
Cited by 2 publications
(10 citation statements)
references
References 24 publications
(29 reference statements)
0
10
0
Order By: Relevance
“…Leaf positioning accuracy affects treatment plan delivery, and inaccurate leaf positioning can cause large dose deviations [19]. We found the leaf positioning accuracy and reproducibility of the αMLC to be <1 mm.…”
Section: Discussionmentioning
confidence: 75%
See 2 more Smart Citations
“…Leaf positioning accuracy affects treatment plan delivery, and inaccurate leaf positioning can cause large dose deviations [19]. We found the leaf positioning accuracy and reproducibility of the αMLC to be <1 mm.…”
Section: Discussionmentioning
confidence: 75%
“…The dual-layer leaf transmission of the αMLC is signi cantly reduced compared to its single-layer. The Elekta Agility MLC [6], the Varian Millennium TM 120-leaf MLC [14,15], the Cobalt-60-based ViewRay MRIdian MLC [7], and Euromechanics PMLC [19] have a measured average leaf transmission of 0.35%, 1.36%, 0.153%, and 1.79%, respectively. The maximum leaf transmission of the above MLC systems is 0.44%, 1.8%, 0.35 %, and 4.1%, respectively.…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…18,19 Generally, a good MLC should possess high leaf positioning accuracy, low leaf transmission, small penumbra, and the ability to accurately create various clinically required complex fields. 6,10,17,20 VenusX is a medical linear accelerator (linac) developed by LinaTech (LinaTech LLC, Sunnyvale, CA), a medical company. Recently, VenusX has been approved by the China Food and Drug Administration and is being used in clinics.…”
Section: Introductionmentioning
confidence: 99%
“…Unfortunately, the application of single‐layer MLC in radiotherapy has many limitations. Radiation leakage increases the dose both inside and outside the target volume, which is undesirable 17 . Additionally, the projection width of the traditional single‐layer MLC on the isocenter plane typically ranges from 0.25 to 1.0 cm.…”
Section: Introductionmentioning
confidence: 99%