The present study examines whether short measurement time and noise filter processing in an ultra-high performance liquid chromatography with ultraviolet detection (UHPLC-UV) contribute to limitations for repeatability assessment based on the ISO 11843 part 7 (ISO 11843-7), which can stochastically provide a measurement standard deviation (SD) caused by baseline noise (SB). In this study, ergosterol was used as an example in UHPLC-UV analysis. From the results of power spectrum analysis of baseline noise, 1024 consecutive digital data points provided a suitable SB. Thus, it was found that an SB can be obtained from about 1 min of baseline when a chromatogram was recorded at sampling rate of 20 points s -1 in the present UHPLC-UV system. The relative SDs (RSDs) of the peak area obtained by the ISO 11843-7 were within 95% of the confidential intervals of the RSDs obtained by repetitive measurements, indicating the ISO 11843-7 is applicable to estimate repeatability in a UHPLC-UV system. In a similar way, we found that the RSD of the peak area obtained from a chromatogram with noise filter processing in UHPLC-UV could also be estimated by the ISO 11843-7. In conclusion, we experimentally demonstrate that short measurement time and noise filter processing are not limitations for repeatability assessment based on the ISO 11843-7.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.