Off-axis energy variation is rather small in unflattened beams and less than half the one for flattened beams. Thus, ignoring the effect of off-axis energy variation for dose calculations in unflattened beams can be clinically justified.
Multileaf collimators (MLCs) need to be characterized accurately in treatment planning systems to facilitate accurate intensity‐modulated radiation therapy (IMRT) and volumetric‐modulated arc therapy (VMAT). The aim of this study was to examine the use of MapCHECK 2 and ArcCHECK diode arrays for optimizing MLC parameters in Monaco X‐ray voxel Monte Carlo (XVMC) dose calculation algorithm. A series of radiation test beams designed to evaluate MLC model parameters were delivered to MapCHECK 2, ArcCHECK, and EBT3 Gafchromic film for comparison. Initial comparison of the calculated and ArcCHECK‐measured dose distributions revealed it was unclear how to change the MLC parameters to gain agreement. This ambiguity arose due to an insufficient sampling of the test field dose distributions and unexpected discrepancies in the open parts of some test fields. Consequently, the XVMC MLC parameters were optimized based on MapCHECK 2 measurements. Gafchromic EBT3 film was used to verify the accuracy of MapCHECK 2 measured dose distributions. It was found that adjustment of the MLC parameters from their default values resulted in improved global gamma analysis pass rates for MapCHECK 2 measurements versus calculated dose. The lowest pass rate of any MLC‐modulated test beam improved from 68.5% to 93.5% with 3% and 2 mm gamma criteria. Given the close agreement of the optimized model to both MapCHECK 2 and film, the optimized model was used as a benchmark to highlight the relatively large discrepancies in some of the test field dose distributions found with ArcCHECK. Comparison between the optimized model‐calculated dose and ArcCHECK‐measured dose resulted in global gamma pass rates which ranged from 70.0%–97.9% for gamma criteria of 3% and 2 mm. The simple square fields yielded high pass rates. The lower gamma pass rates were attributed to the ArcCHECK overestimating the dose in‐field for the rectangular test fields whose long axis was parallel to the long axis of the ArcCHECK. Considering ArcCHECK measurement issues and the lower gamma pass rates for the MLC‐modulated test beams, it was concluded that MapCHECK 2 was a more suitable detector than ArcCHECK for the optimization process.PACS number(s): 87.55.Qr
The aim of this study was to investigate the dosimetric performance of a novel optical fiber sensor for use in external beam radiation therapy. Repeatability and reproducibility of the output signal, linearity, dose rate and dose per pulse dependence were evaluated. Angular dependence was investigated in the axial and azimuthal planes. The percentage depth dose and lateral dose profiles were measured using the optical fiber sensor system and compared to commercially available detectors such as Exradin W1 plastic scintillator and a PTW-microdiamond detector. The result of this study show that the optical fiber sensor system has good repeatability and reproducibility of the output signal with a maximum deviation of 0.17% and 1.00%, respectively. The system also showed an excellent linearity with dose, and its signal was independent of dose rate. However, the system showed a strong dependence on dose per pulse with 27% deviation from the W1 result at the highest dose per pulse value that was achieved at 75 cm source to surface distance. The system also showed an angular dependence when the incident beam was in the azimuthal plane due to the geometry of the scintillator at the tip of the fiber. The optical fiber sensor overresponded when measuring percentage depth dose curves and lateral dose profiles due in part to the sensitivity of the scintillating material (Gd2O2S:Tb) to low energy scattered radiation. However, further investigation is needed to quantify the overall contribution of Cerenkov radiation to the over-response of the optical fiber sensor.
The characteristics of an Elekta amorphous silicon (a-Si) electronic portal imaging device (EPID) in response to a 6 MV photon beam generated without a flattening filter, an unflattened beam, have been determined. The characteristics were then compared to those for a conventional photon beam generated with a flattening filter in the beam, a flattened beam, in order to determine the suitability of an a-Si EPID for transit dosimetry. The response of the EPID to the unflattened beam increased by 7.3% compared to the flattened beam, and copper buildup of 3 mm reduces the variation in the EPID response over air gaps ranging from 60 to 40 cm to within 2.5%. The scattering properties of the EPID with changing field size for the unflattened beam agree with those measured for a flattened beam to within 2%. Due to the minimal variation in the energy spectrum of the unflattened beam with the distance from the central axis, it was expected and experimentally found that the profile shape of the unflattened beam changes minimally with increasing phantom thickness. For an unflattened beam, EPID measured profiles with and without a phantom in the beam agree to within 2% using confidence limits. The difference between EPID and ionization chamber profiles measured at a depth of 5 cm in water is reduced compared to a flattened beam and remains unchanged with increasing phantom thickness. A difference of 4% was found between EPID profiles and the corresponding profiles measured with an ionization chamber measured in water over a range of phantom thickness. A calibration procedure was developed to convert EPID images to the equivalent absolute dose in water, at the EPID plane. A gamma evaluation was performed comparing the calibrated EPID images to dose measured with an ionization chamber array for rectangular fields and an IMRT segment. The fields were situated on axis and at 5 cm off axis with and without a 25 cm thick phantom in the beam. The gamma evaluation criteria of 3% and 3 mm were met within the field, for all fields examined. This study concludes that a-Si EPIDs are suitable dosimeters for IMRT treatments using unflattened photon beams with the advantage that the characteristics of the unflattened beam result in a reduction in the number of measurements necessary to calibrate an a-Si EPID as a transit dosimeter.
This note outlines an improved method of calculating dose per monitor unit values for small electron fields using Khan's lateral build-up ratio (LBR). This modified method obtains the LBR directly from the ratio of measured, surface normalized, electron beam percentage depth dose curves. The LBR calculated using this modified method more accurately accounts for the change in lateral scatter with decreasing field size. The LBR is used along with Khan's dose per monitor unit formula to calculate dose per monitor unit values for a set of small fields. These calculated dose per monitor unit values are compared to measured values to within 3.5% for all circular fields and electron energies examined. The modified method was further tested using a small triangular field. A maximum difference of 4.8% was found.
Several instruments have been used to measure absorbed radiation dose under non-electronic equilibrium conditions, such as in the build-up region or near the interface between two different media, including the surface. Many of these detectors are discussed in this paper. A common method of measuring the absorbed dose distribution and electron contamination in the build-up region of high-energy beams for radiation therapy is by means of parallel-plate ionisation chambers. Thermoluminescent dosimeters (TLDs), diodes and radiographic film have also been used to obtain surface dose measurements. The diamond detector was used recently by the author in an investigation on the effects of beam-modifying devices on skin dose and it is also described in this report.
Aim: To determine the energy and dose dependence of GafChromic EBT3-V3 film over an energy range 0.2 mm Al HVL to 6 MV. Background: The decay scheme of a brachytherapy source may be complex and the spectrum of energy can be wide. LiF TLDs are the golden standard recommended for dosimetric measures in brachytherapy, for their energy independence, but TLDs could be not available in some centres. An alternative way to perform dose measurements is to use GafChromic films, but they show energy dependence. Methods and materials: Films have been irradiated at increasing dose with three different beams: 6 MV beam, TPR20, 10 = (0.684 ± 0.01), HVL = (2.00 ± 0.01)mmAl and HVL = (0.20 ± 0.01)mmAl. Calibration curves were generated using the same dose range (0cGy to 850cGy) for the three energies. Using the 6 MV calibration curve as reference, the film response in terms of net optical density (OD) was evaluated. Results: The difference in the calibration curve obtained by irradiating the film with 6 MV and 2 mm Al HVL energy beams is less than 3 %, within the calibration uncertainty, in the dose range 500-850cGy. The OD of EBT3-V3 film is significantly lower at 0.2 mmAl HVL compared to 6 MV, showing differences up to 25 %. Conclusion: Within the range 6 MV-2 mm Al HVL and dose higher than 500cGy, GafChromic EBT3-V3 films are energy independent. In this dose range, films can be calibrated in a simple geometry, using a 6 MV Linac beam, and can be used for brachytherapy sources dose measures. The use of EBT3 films can be extended to reference dosimetry in Ir-192 clinical brachytherapy.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
hi@scite.ai
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.