Energy and dose dependence of GafChromic EBT3-V3 film across a wide energy range

Abstract: Aim: To determine the energy and dose dependence of GafChromic EBT3-V3 film over an energy range 0.2 mm Al HVL to 6 MV. Background: The decay scheme of a brachytherapy source may be complex and the spectrum of energy can be wide. LiF TLDs are the golden standard recommended for dosimetric measures in brachytherapy, for their energy independence, but TLDs could be not available in some centres. An alternative way to perform dose measurements is to use GafChromic films, but they show energy dependence. Methods a… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
2

Citation Types

0
7
0

Year Published

2020
2020
2023
2023

Publication Types

Select...
6

Relationship

1
5

Authors

Journals

citations
Cited by 8 publications
(7 citation statements)
references
References 5 publications
0
7
0
Order By: Relevance
“…In order to calibrate GafChromic EBT3-V3 films, an energy independence check was performed as published by Bassi et al [28]. Film was irradiated at increasing doses with three different beams: 6 MV beam from Elekta Precise Linac, TPR 20,10 = 0.684 ± 0.01, HVL = 2.00 ± 0.01 mm Al and HVL = 0.20 ± 0.01 mm Al, from Xstrahl Gulmay Orthovoltage unit.…”
Section: Gafchromic Filmmentioning
confidence: 99%
See 1 more Smart Citation
“…In order to calibrate GafChromic EBT3-V3 films, an energy independence check was performed as published by Bassi et al [28]. Film was irradiated at increasing doses with three different beams: 6 MV beam from Elekta Precise Linac, TPR 20,10 = 0.684 ± 0.01, HVL = 2.00 ± 0.01 mm Al and HVL = 0.20 ± 0.01 mm Al, from Xstrahl Gulmay Orthovoltage unit.…”
Section: Gafchromic Filmmentioning
confidence: 99%
“…Hence, EBT3-V3 films are suitable dose detectors when the dose response curve is measured within the range 6 MV and 2 mm Al HVL. 6 MV calibration curve could be used within this range, with 3%, k = 2, uncertainty [28].…”
Section: Gafchromic Filmmentioning
confidence: 99%
“…Based on the literature, the most frequently reported dosimetry systems in the determination of the dosimetric function of 192 Ir source are film dosimetry and thermoluminescent dosimeter (TLD) (Ayoobian et al, 2016;Bassi et al, 2020;DeWerd et al, 2014;Sellakumar et al, 2009;Uniyal et al, 2011;Wu et al, 2014). The dosimetric functions of 192 Ir studied by Sellakumar et al (2009) were measured the Λ , , and F(r, θ) for HDR 192 Ir source using Gafchromic® EBT films.…”
Section: Introductionmentioning
confidence: 99%
“…A recent study by Bassi et al (2020) reported, EBT3 film was energy independent and can be utilised for brachytherapy source dose monitoring despite being calibrated with a 6MV photon beam and could be expanded to be applied in clinical dosimetry brachytherapy. Besides film dosimetry, TLDs also was recommended for dosimetric measurement in brachytherapy because of their excellent sensitivity, miniature, flat energy response, and energy independence.…”
Section: Introductionmentioning
confidence: 99%
See 1 more Smart Citation