BackgroundPlan adaptation during the course of (chemo)radiotherapy of H&N cancer requires repeat CT scanning to capture anatomy changes such as parotid gland shrinkage. Hydration, applied to prevent nephrotoxicity from cisplatin, could temporarily alter the hydrogen balance and hence the captured anatomy. The aim of this study was to determine geometric changes of parotid glands as function of hydration during chemoradiotherapy compared to a control group treated with radiotherapy only.MethodsThis study included an experimental group (n = 19) receiving chemoradiotherapy, and a control group (n = 19) receiving radiotherapy only. Chemoradiotherapy patients received cisplatin with 9 l of saline solution during hydration in the first, fourth and seventh week. The delineations of the parotid glands on the planning CT scan were automatically propagated to Cone Beam CT scans using deformable image registration. Relative volume and position of the parotid glands were determined at the second chemotherapy cycle (week four) and at fraction 35.ResultsWhen saline solution was administrated, the volume temporarily increased on the first day (7.2 %, p < 0.001), second day (10.8 %, p < 0.001) and third day (7.0 %, p = 0.016). The gland positions shifted lateral, the distance between glands increased on the first day with 1.5 mm (p < 0.001), on the second day 2.2 mm (p < 0.001). At fraction 35, with both groups the mean shrinkage was 24 % ± 11 % (1SD) and the mean medial distance between the parotid glands decreased by 0.47 cm ± 0.27 cm.ConclusionsHydration significantly modulates parotid gland geometry. Unless, in the context of adaptive RT, a repeat CT scan is timed during a chemotherapy cycle, these effects are of minor clinical relevance.
Purpose:
To examine CT metal image artifact from a novel direction‐modulated brachytherapy (DMBT) tandem applicator (95% tungsten) for cervical cancer using a commercially available orthopedic metal artifact reduction (O‐MAR) algorithm. Comparison to a conventional stainless steel applicator is also performed.
Methods:
Each applicator was placed in a water‐filled phantom resembling the female pelvis and scanned in a Philips Brilliance 16‐slice CT scanner using two pelvis protocols: a typical clinical protocol (120kVp, 16×0.75mm collimation, 0.692 pitch, 1.0s rotation, 350mm field of view (FOV), 600mAs, 1.5mm slices) and a protocol with a higher kVp and mAs setting useful for larger patients (140kVp, 16×0.75mm collimation, 0.688 pitch, 1.5s rotation, 350mm FOV, 870mAs, 1.5mm slices). Images of each tandem were acquired with and without the application of the O‐MAR algorithm. Baseline scans of the phantom (no applicator) were also collected. CT numbers were quantified at distances from 5 to 30 mm away from the applicator's edge (in increments of 5mm) using measurements at eight angles around the applicator, on three consecutive slices.
Results:
While the presence of both applicators degraded image quality, the DMBT applicator resulted in larger streaking artifacts and dark areas in the image compared to the stainless steel applicator. Application of the O‐MAR algorithm improved all acquired images, both visually and quantitatively. The use of low and high kVp and mAs settings (120 kVp/600mAs and 140 kVp/870mAs) in conjunction with the O‐MAR algorithm lead to similar CT numbers in the vicinity of the applicator and a similar reduction of the induced metal artifact.
Conclusion:
This work indicated that metal artifacts induced by the DMBT and the stainless steel applicator are greatly reduced when using the O‐MAR algorithm, leading to better quality phantom images. The use of a high dose protocol provided similar improvements in metal artifacts compared to the clinical protocol.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.