This paper presents a numerical study of the effects of inlet supply starvation on film thickness in EHL point contacts. Generally this problem is treated using the position of the inlet meniscus as the governing parameter; however, it is difficult to measure this in real applications. Thus, in this paper an alternative approach is adopted whereby the amount of oil present on the surfaces is used to define the degree of starvation. It is this property which determines both meniscus position and film thickness reduction. The effect of subsequent overrollings on film thickness decay can also be evaluated. In the simplest case a constant lubricant inlet film thickness in the Y direction is assumed and the film thickness distribution is computed as a function of the oil available. This yields an equation predicting the film thickness reduction, with respect to the fully flooded value, from the amount of lubricant initially available on the surface, as a function of the number of overrollings n. However, the constant inlet film thickness does not give a realistic description of starvation for all conditions. Some experimental studies show that the combination of side flow and replenishment action can generate large differences in local oil supply and that the side reservoirs play an important role in this replenishment mechanism. Thus the contact centre can be fully starved whilst the contact sides remain well lubricated. In these cases, a complete analysis with a realistic inlet distribution has been carried out and the numerical results agree well with experimental findings.
The revision rate of large head metal-on-metal and resurfacing hips are significantly
higher than conventional total hip replacements. The revision of these components has been
linked to high wear caused by edge loading; which occurs when the head–cup contact patch
extends over the cup rim. There are two current explanations for this; first, there is
loss of entrainment of synovial fluid resulting in breakdown of the lubricating film and
second, edge loading results in a large local increase in contact pressure and consequent
film thickness reduction at the cup rim, which causes an increase in wear.This paper develops a method to calculate the distance between the joint reaction force
vector and the cup rim – the contact patch centre to rim (CPCR) distance. However, the
critical distance for the risk of edge loading is the distance from the contact patch edge
to rim (CPER) distance. An analysis of explanted hip components, divided into edge worn
and non-edge-worn components showed that there was no statistical difference in CPCR
values, but the CPER value was significantly lower for edge worn hips.Low clearance hips, which have a more conformal contact, have a larger diameter contact
patch and thus are more at risk of edge loading for similarly positioned hips.
Lubricant films were measured for a series of bovine serum and protein containing (albumin, globulin) saline solutions for CoCrMo femoral component sliding against a glass disc. Central film thickness was measured by optical interferometry as a function of time (constant mean speed: 0 and 10 mm/s) and variable mean speed (0-50 mm/s). The effect of load (5-20 N) on film thickness was also studied. The development of the wear scar on the CoCrMo surface was monitored by measuring the width of the contact zone during the film thickness tests. The results showed film thickness increased with time for both the static and sliding tests. Films formed in the static, loaded test were typically in the range of 3-40 nm. The globulin containing solutions formed the thickest films. In the sliding tests a wear scar rapidly formed on the implant component for the bovine serum and albumin fluids, negligible wear was observed for the globulin solutions. Film thickness increased with sliding time for all test solutions and was much greater than predicted by isoviscous EHL models. The film increase was found to correlate with increasing wear scar size and thus decreasing contact pressure. A new lubricating mechanism is proposed whereby during sliding the fluid undergoes bulk phase separation rheology, so that an elevated protein phase forms in the inlet zone. This protein phase is a high-viscosity biphasic matrix, which is periodically entrained into the contact forming a thick protective hydro-gel film. One of the main findings of this study is that film thickness was very sensitive to load; to a much greater extent than predicted by EHL models. Thus film formation in MoM hip joints is very susceptible to high contact pressures which might be due to implant misalignment and edge-loading.
This study compared component wear rates and pre-revision blood metal ions levels in two groups of failed metal-on-metal hip arthroplasties: hip resurfacing and modular total hip replacement (THR). There was no significant difference in the median rate of linear wear between the groups for both acetabular (p = 0.4633) and femoral (p = 0.0872) components. There was also no significant difference in the median linear wear rates when failed hip resurfacing and modular THR hips of the same type (ASR and Birmingham hip resurfacing (BHR)) were compared. Unlike other studies of well-functioning hips, there was no significant difference in pre-revision blood metal ion levels between hip resurfacing and modular THR. Edge loading was common in both groups, but more common in the resurfacing group (67%) than in the modular group (57%). However, this was not significant (p = 0.3479). We attribute this difference to retention of the neck in resurfacing of the hip, leading to impingement-type edge loading. This was supported by visual evidence of impingement on the femur. These findings show that failed metal-on-metal hip resurfacing and modular THRs have similar component wear rates and are both associated with raised pre-revision blood levels of metal ions.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.