Background Pseudotumors are sterile inflammatory lesions found in the soft tissues surrounding metal-on-metal (MOM) and metal-on-polyethylene hip arthroplasties. In patients with MOM hip arthroplasties, pseudotumors are thought to represent an adverse reaction to metal wear debris. However, the pathogenesis of these lesions remains unclear. Currently, there is inconsistent evidence regarding the influence of adverse cup position and increased wear in the formation of pseudotumors. Questions/purposes We therefore determined whether pseudotumor formation was associated with (1) adverse cup position, (2) raised metal ion levels, and (3) increased wear rates of the retrieved components. Methods We retrospectively reviewed all 352 patients for whom we had retrieved specimens from revisions of a current-generation MOM hip prosthesis between February 2008 and September 2010; of these, 105 met our inclusion criteria. We used multivariate logistic regression analysis to compare acetabular orientation, metal ion levels before revision, and component wear rates between patients with (n = 72) and without (n = 33) pseudotumors, according to findings on metal artifact reduction sequence MRI. Results The proportion of patients demonstrating evidence of a pseudotumor in well-positioned hips was similar to those with adverse cup positions (67% and 66%, respectively). Patients revised with pseudotumors had similar whole-blood metal ion levels and component wear rates to those who were not revised. Conclusions Pseudotumors were not associated with increased wear or metal ion levels, suggesting patient susceptibility is likely to be more important.
It has been speculated that material loss, either as corrosion or wear, at the head-stem taper junction is implicated in the high revision rates reported for metal-on-metal total hip replacements. We measured the volume of material loss from the taper and bearing surfaces of retrieved devices, and investigated the associations with blood metal ion levels and the diagnosis of a cystic or solid pseudotumor. The median volumes of material lost from the female and male taper surfaces were 2.0 and 0.29 mm 3 , respectively, while the median volumes of wear from the cup and head bearing surfaces were 1.94 and 3.44 mm 3 , respectively. Material loss from the female taper was similar to that from the acetabular bearing surface (p ¼ 0.55), but significantly less than that from the femoral bearing surface (p < 0.001). Material loss from the male taper was less than that from both bearing surfaces (p < 0.001). Multivariable analysis demonstrated no significant correlations between the volume of material lost from the taper surfaces and either blood cobalt or chromium ions, or the presence of pseudotumor. While a substantial volume of material is lost at the taper junction, the clinical significance of this debris remains unclear. ß
This study compared component wear rates and pre-revision blood metal ions levels in two groups of failed metal-on-metal hip arthroplasties: hip resurfacing and modular total hip replacement (THR). There was no significant difference in the median rate of linear wear between the groups for both acetabular (p = 0.4633) and femoral (p = 0.0872) components. There was also no significant difference in the median linear wear rates when failed hip resurfacing and modular THR hips of the same type (ASR and Birmingham hip resurfacing (BHR)) were compared. Unlike other studies of well-functioning hips, there was no significant difference in pre-revision blood metal ion levels between hip resurfacing and modular THR. Edge loading was common in both groups, but more common in the resurfacing group (67%) than in the modular group (57%). However, this was not significant (p = 0.3479). We attribute this difference to retention of the neck in resurfacing of the hip, leading to impingement-type edge loading. This was supported by visual evidence of impingement on the femur. These findings show that failed metal-on-metal hip resurfacing and modular THRs have similar component wear rates and are both associated with raised pre-revision blood levels of metal ions.
The Articular Surface Replacement (ASR) hip resurfacing arthroplasty has a failure rate of 12.0% at five years, compared with 4.3% for the Birmingham Hip Resurfacing (BHR). We analysed 66 ASR and 64 BHR explanted metal-on-metal hip replacements with the aim of understanding their mechanisms of failure. We measured the linear wear rates of the acetabular and femoral components and analysed the clinical cause of failure, pre-revision blood metal ion levels and orientation of the acetabular component. There was no significant difference in metal ion levels (chromium, p = 0.82; cobalt, p = 0.40) or head wear rate (p = 0.14) between the two groups. The ASR had a significantly increased rate of wear of the acetabular component (p = 0.03) and a significantly increased occurrence of edge loading (p < 0.005), which can be attributed to differences in design between the ASR and BHR. The effects of differences in design on the in vivo wear rates are discussed: these may provide an explanation as to why the ASR is more sensitive to suboptimal positioning than the BHR.
Suboptimal component position and design are thought to lead to edge wear and raised blood metal ion levels in metal-onmetal hip resurfacing (MOM-HR). These factors are thought to influence the "contact patch to rim distance" (CPRD), and calculation of this distance may improve prediction of wear and blood metal ion levels. We measured blood cobalt and chromium ion levels and the wear rates of the bearing surfaces in 165 MOM-HR retrieval cases. We then determined the contribution and effect sizes of cup inclination and version angles, component size and design, and CPRD (calculated from case-specific data) on blood metal ion levels and component wear rates. Acetabular orientation explained between 16.3% and 28.5% of the variation in wear rates and metal ion levels, whereas component size and design explained between 7.3% and 21.8% of the variability. In comparison, CPRD explained up to 67.7% of the variability, significantly greater than any other variable (all p < 0.0001). CPRD is a good predictor of wear and improves our understanding of wear performance and the mechanisms leading to edge loading. Keywords: metal-on-metal; hip resurfacing; retrievals; wear; metal ions High revision rates have been widely reported for metal-on-metal hip resurfacing (MOM-HR), 1 and subsequently the use of MOM has become a major public health concern subject to widespread coverage in both the scientific and lay press. High revision rates have been partly attributed to the release of metal debris from the bearing surfaces. High wear, occurring as a result of edge loading, has been associated with devastating inflammatory soft tissue reactions. 2-4These patients have poorer outcomes after revision surgery. 5Previous studies reported a significant association between cup inclination angle and both wear rates and blood metal ion levels.6-8 However, the correlations with cup inclination are only weakly positive and therefore can not alone explain all the variation in wear rates and ion levels. 6,8 Indeed, edge loading in MOM-HR is likely attributable not only to component position, but also component design 9,10 and patient activity patterns. 11The Articular Surface Replacement (ASR; DePuy Orthopaedics) is widely recognized as having the poorest clinical and wear performance of the contemporary MOM-HR designs. 1,12,13 Design features such as the reduced "arc of cover" (the angle subtended by the articular surface of the cup) and head-cup clearance are thought to increase the likelihood of edge loading and high wear.12,13 An overview of the design variations among manufacturers is shown in Table 1.Many of the variables that affect wear, both surgical and design, likely do so by moving the contact area between the femoral and acetabular bearing surfaces closer to the cup rim. This distance has been described in the literature , and in this study we have termed it the 'contact patch to rim distance' (CPRD). As shown by the two-dimensional schematic (Fig. 1), a reduction in this distance increase the likelihood of edge loading a...
BackgroundEarly transfusion of packed red blood cells (PRBC) has been associated with improved survival in patients with haemorrhagic shock. This study aims to describe the characteristics of patients receiving pre-hospital blood transfusion and evaluate their subsequent need for in-hospital transfusion and surgery.MethodsThe decision to administer a pre-hospital PRBC transfusion was based on clinical judgment. All patients transfused pre-hospital PRBC between February 2013 and December 2014 were included. Pre-hospital and in-hospital records were retrospectively reviewed.ResultsOne hundred forty-seven patients were included. 142 patients had traumatic injuries and 5 patients had haemorrhagic shock from a medical origin. Median Injury Severity Score was 30. 90% of patients receiving PRBC had an ISS of >15. Patients received a mean of 2.4(±1.1) units of PRBC in the pre-hospital phase. Median time from initial emergency call to hospital arrival was 114 min (IQR 103–140). There was significant improvement in systolic (p < 0.001), diastolic (p < 0.001) and mean arterial pressures (p < 0.001) with PRBC transfusion but there was no difference in HR (p = 0.961). Patients received PRBC significantly faster in the field than waiting until hospital arrival. At the receiving hospital 57% required an urgent surgical or interventional radiology procedure. At hospital arrival, patients had a mean lactate of 5.4(±4.4) mmol/L, pH of 6.9(±1.3) and base deficit of −8.1(±6.7). Mean initial serum adjusted calcium was 2.26(±0.29) mmol/L. 89% received further blood products in hospital. No transfusion complications or significant incidents occurred and 100% traceability was achieved.DiscussionPre-hospital transfusion of packed red cells has the potential to improvde outcome for trauma patients with major haemorrhage. The pre-hospital time for trauma patients can be several hours, suggesting transfusion needs to start in the pre-hospital phase. Hospital transfusion research suggests a 1:1 ratio of packed red blood cells to plasma improves outcome and further research into pre-hospital adoption of this strategy is needed.ConclusionPre-hospital PRBC transfusion significantly reduces the time to transfusion for major trauma patients with suspected major haemorrhage. The majority of patients receiving pre-hospital PRBC were severely injured and required further transfusion in hospital. Further research is warranted to determine which patients are most likely to have outcome benefit from pre-hospital blood products and what triggers should be used for pre-hospital transfusion.
Multivariate analysis of nine factors found that edge-loading was the most important predictor of wear rate and occurred in two-thirds of failed metal-on-metal hip replacements. The majority did not have excessive cup inclination angles: 68% had an inclination angle of ≤55°. This finding, together with the relatively low median wear rate of the components in our study, suggests that cup position and/or wear rate may not be the only outcome related to failure of metal-on-metal hip replacements.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.