Lateral flow assays (LFAs) are a widely-used point-of care diagnostic format, but suffer from limited analytical sensitivity, especially when read by eye. It has recently been reported that LFA performance can be improved by using magnetic reporter particles and an external magnetic field applied at the test line. The mechanism of sensitivity/performance enhancement was suggested to be concentration/retardation of reporter particles at the test line. Here we demonstrate an additional mechanism of particle relocation where reporter particles from the lower depths of the translucent LFA strip relocate to more-visible locations nearer to the top surface, producing a more visible signal. With a magnetic field we observed an improvement in sensitivity of human chorionic gonadotropin (hCG) detection from 1.25 ng/mL to 0.31 ng/mL. We also observed an increase of the color intensity per particle in test lines when the magnetic field was present.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.