The 8 and 27 keV monoenergetic neutron calibration fields have been developed by using (45)Sc(p, n)(45)Ti reaction. Protons from a 4-MV Pelletron accelerator are used to bombard a thin scandium target evaporated onto a platinum disc. The proton energies are finely adjusted to the resonance to generate the 8 and 27 keV neutrons by applying a high voltage to the target assemblies. The neutron energies were measured using the time-of-flight method with a lithium glass scintillation detector. The neutron fluences at a calibration point located at 50 cm from the target were evaluated using Bonner spheres. A long counter was placed at 2.2 m from the target and at 60 degrees to the direction of the proton beam in order to monitor the fluence at the calibration point. Fluence and dose equivalent rates at the calibration point are sufficient to calibrate many types of the neutron survey metres.
Monoenergetic neutron calibration fields of 144, 565 keV and 5.0 MeV have been developed at the Facility of Radiation Standards of JAERI using a 4 MV Pelletron accelerator. The 7Li(p,n)7Be and 2H(d,n)3He reactions are employed for neutron production. The neutron energy was measured by the time-of-flight method with a liquid scintillation detector and calculated with the MCNP-ANT code. A long counter is employed as a neutron monitor because of the flat response. The monitor is set up where the influence of inscattered neutrons from devices and their supporting materials at a calibration point is as small as possible. The calibration coefficients from the monitor counts to the neutron fluence at a calibration point were obtained from the reference fluence measured with the transfer instrument of the primary standard laboratory (AIST), a 24.13 cm phi Bonner sphere counter. The traceability of the fields to AIST was established through the calibration.
A dose evaluation method for neutrons in the energy range of a few MeV to 100 MeV has been developed using a spectrum weight function (G-function), which is applied to an organic liquid scintillator of 12.7 cm in diameter and 12.7 cm in length. The G-function that converts the pulse height spectrum of the scintillator into the ambient dose equivalent, H*(10), was calculated by an unfolding method using successive approximation of the response function of the scintillator and the ambient dose equivalent per unit neutron fluence (H*(10) conversion coefficients) of ICRP 74. To verify the response function of the scintillator and the value of H*(10) evaluated by the G-function. pulse height spectra of the scintillator were measured in some different neutron fields, which have continuous energy, monoenergetic and quasi-monoenergetic spectra. Values of H*(10) estimated using the G-function and pulse height spectra of the scintillator were compared with those calculated using neutron energy spectra. These doses agreed with each other. From the results, it was concluded that H*(10) can be evaluated directly from the pulse height spectrum of the scintillator by applying the G-function proposed in this study.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.