2007
DOI: 10.1093/rpd/ncm004
|View full text |Cite
|
Sign up to set email alerts
|

Construction of monoenergetic neutron calibration fields using 45Sc(p, n)45Ti reaction at JAEA

Abstract: The 8 and 27 keV monoenergetic neutron calibration fields have been developed by using (45)Sc(p, n)(45)Ti reaction. Protons from a 4-MV Pelletron accelerator are used to bombard a thin scandium target evaporated onto a platinum disc. The proton energies are finely adjusted to the resonance to generate the 8 and 27 keV neutrons by applying a high voltage to the target assemblies. The neutron energies were measured using the time-of-flight method with a lithium glass scintillation detector. The neutron fluences … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
15
0
1

Year Published

2008
2008
2023
2023

Publication Types

Select...
7

Relationship

1
6

Authors

Journals

citations
Cited by 18 publications
(16 citation statements)
references
References 3 publications
0
15
0
1
Order By: Relevance
“…The calibration coefficients K were determined for every energy point by measuring the reference neutron fluence at the calibration point using neutron detectors such as a Bonner sphere, a proton recoil proportional counter and a semiconductor detector with a polyethylene convertor [1][2][3]. The correction factor S is determined for every calibration by measuring the ratio of monitor counts with and without the instruments and their supporting materials.…”
Section: Determination Of Neutron Fluencementioning
confidence: 99%
See 2 more Smart Citations
“…The calibration coefficients K were determined for every energy point by measuring the reference neutron fluence at the calibration point using neutron detectors such as a Bonner sphere, a proton recoil proportional counter and a semiconductor detector with a polyethylene convertor [1][2][3]. The correction factor S is determined for every calibration by measuring the ratio of monitor counts with and without the instruments and their supporting materials.…”
Section: Determination Of Neutron Fluencementioning
confidence: 99%
“…For the 8 and 27 keV fields, the aimed neutron can be produced by adjusting the proton energy to the resonance peak of the 45 Sc(p,n) 45 Ti reaction with a target voltage control system [2].…”
Section: Energy Monitoringmentioning
confidence: 99%
See 1 more Smart Citation
“…The position among several positions available from the arrangement in the room was determined to make the contribution of inscattered neutrons from the devices to be calibrated and their supporting materials small as possible. Calibration coefficients K LC , which are used to calculate the reference neutron fluence at a calibration point from the counts of the LC N LC , were determined from the fluence measured by using the detectors mentioned in the following sections 1,2) .…”
Section: Outline Of the Facility At Frsmentioning
confidence: 99%
“…Therefore, such monoenergetic neutron calibration fields below 20 MeV have been developed as the secondary standard fields using a 4 MV Van-de-Graaf accelerator at the Facility of Radiation Standards (FRS) of the Japan Atomic Energy Agency (JAEA) 1,2) . The calibration fields of 8, 144, 250, 565 keV, 5.0 and 14.8 MeV neutrons are available at FRS (Fig.…”
Section: Introductionmentioning