Cannulated screws, along with guide wires, are typically used for surgical fracture treatment in cancellous bone. Breakage or bending deformation of the guide wire is a clinical concern. Mechanically superior guide wires made of Co-Cr alloys such as MP35N and L605 may reduce the occurrence of mechanical failures when used in combination with conventional (316L stainless steel) cannulated screws. However the possibility of galvanic or crevice corrosion and adverse tissue reaction, exists when using dissimilar materials, particularly in the event that a guide wire breaks, and remains in situ. Therefore, we designed an experiment to determine the tissue reaction to such an in vivo environment. Implant devices were designed to replicate a clinical situation where dissimilar metals can form a galvanic couple. Histological and SEM analyses were used to evaluate tissue response and corrosion of the implants. In this experiment, no adverse in vivo effects were detected from the use of dissimilar materials in a model of a broken guide wire in a cannulated screw.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.