Purpose
To develop a highly efficient magnetic field gradient coil for head imaging that achieves 200 mT/m and 500 T/m/s on each axis using a standard 1 MVA gradient driver in clinical whole‐body 3.0T MR magnet.
Methods
A 42‐cm inner diameter head‐gradient used the available 89‐ to 91‐cm warm bore space in a whole‐body 3.0T magnet by increasing the radial separation between the primary and the shield coil windings to 18.6 cm. This required the removal of the standard whole‐body gradient and radiofrequency coils. To achieve a coil efficiency ~4× that of whole‐body gradients, a double‐layer primary coil design with asymmetric x‐y axes, and symmetric z‐axis was used. The use of all‐hollow conductor with direct fluid cooling of the gradient coil enabled ≥50 kW of total heat dissipation.
Results
This design achieved a coil efficiency of 0.32 mT/m/A, allowing 200 mT/m and 500 T/m/s for a 620 A/1500 V driver. The gradient coil yielded substantially reduced echo spacing, and minimum repetition time and echo time. In high b = 10,000 s/mm2 diffusion, echo time (TE) < 50 ms was achieved (>50% reduction compared with whole‐body gradients). The gradient coil passed the American College of Radiology tests for gradient linearity and distortion, and met acoustic requirements for nonsignificant risk operation.
Conclusions
Ultra‐high gradient coil performance was achieved for head imaging without substantial increases in gradient driver power in a whole‐body 3.0T magnet after removing the standard gradient coil. As such, any clinical whole‐body 3.0T MR system could be upgraded with 3‐4× improvement in gradient performance for brain imaging.
Electrostatic influences on NMR parameters are well accepted. Experimental and computational routes have been long pursued to understand and utilize such Stark effects. However, existing approaches are largely indirect informants on electric fields, and/or are complicated by multiple causal factors in spectroscopic change. We present a system to directly measure quadrupolar Stark effects from an applied electric (E) field. Our apparatus and applications are relevant in two contexts. Each uses a radiofrequency (rf) E field at twice the nuclear Larmor frequency (2omega(0)). The mechanism is a distortion of the E-field gradient tensor that is linear in the amplitude (E(0)) of the rf E field. The first uses 2omega(0) excitation of double-quantum transitions for times similar to T(1) (the longitudinal spin relaxation time). This perturbs the steady state distribution of spin population. Nonlinear analysis versus E(0) can be used to determine the Stark response rate. The second context uses POWER (perturbations observed with enhanced resolution) NMR. Here, coherent, short-time (<
Clinical studies have established a strong benefit from adjuvant mild hyperthermia (HT) to radio- and chemotherapy for many tumor sites, including the head and neck (H&N). The recently developed HYPERcollar allows the application of local radiofrequency HT to tumors in the entire H&N. Treatment quality is optimized using electromagnetic and thermal simulators and, whenever placement risk is tolerable, assessed using invasively placed thermometers. To replace the current invasive procedure, we are investigating whether magnetic resonance (MR) thermometry can be exploited for continuous and 3D thermal dose assessment. In this work, we used our simulation tools to design an MR compatible laboratory prototype applicator. By simulations and measurements, we showed that the redesigned patch antennas are well matched to 50 Ω (S11<-10 dB). Simulations also show that, using 300 W input power, a maximum specific absorption rate (SAR) of 100 W kg(-1) and a temperature increase of 4.5 °C in 6 min is feasible at the center of a cylindrical fat/muscle phantom. Temperature measurements using the MR scanner confirmed the focused heating capabilities and MR compatibility of the setup. We conclude that the laboratory applicator provides the possibility for experimental assessment of the feasibility of hybrid MR-HT in the H&N region. This versatile design allows rigorous analysis of MR thermometry accuracy in increasingly complex phantoms that mimic patients' anatomies and thermodynamic characteristics.
Radiofrequency electric (E) fields oscillating at twice the usual NMR frequency (2ω(0)) can induce double-quantum transitions in quadrupolar nuclei, an NMR Stark effect. Characterization of such is of interest to aid understanding of electrostatic effects in NMR spectra. Calibration of Stark responses to an applied electric field may also be used to assess native fields within molecules and materials. We present high-field (14.1 T), room-temperature NMR experiments to calibrate the 2ω(0) Stark response in crystalline GaAs. This system presents an important test of current techniques and conditions, as historical studies at low field (500-900 mT) and low temperature (77 K) provide a basis for comparison. Our measurements of steady state response reveal the quadrupolar Stark tuning rate for (69)Ga in this material. The value, β(Q) = (11.5 ± 0.1) × 10(12) m(-1), is 3.6 times larger than the most-reliable prior result. In the process, we also uncovered a previously unobserved double-quantum steady state coherence. It appears as a completely separable dispersive signal component in quadrature-detected presaturation spectra versus offset from 2ω(0). The new component may eventually afford an independent route to calibrating β(Q). Finally, we demonstrated exceptional agreement with theory of the orientation-dependent Stark response for rotation of the sample relative to B(0) over a range of 90° and for E-field amplitudes from 30-180 V/cm.
The techniques presented here may be used to simplify calibration experiments that use an interstitial heating device, or any heating method that provides rapid and spatially localised heat distributions. Overall, the experimental verification of the data registration and PRFS thermal coefficient calibration technique provides a useful benchmarking method to maximise MRT accuracy in any similar context.
BackgroundThe study aims to investigate different ground plane segmentation designs of an ultrasound transducer to reduce gradient field induced eddy currents and the associated geometric distortion and temperature map errors in echo-planar imaging (EPI)-based MR thermometry in transcranial magnetic resonance (MR)-guided focused ultrasound (tcMRgFUS).MethodsSix different ground plane segmentations were considered and the efficacy of each in suppressing eddy currents was investigated in silico and in operando. For the latter case, the segmented ground planes were implemented in a transducer mockup model for validation. Robust spoiled gradient (SPGR) echo sequences and multi-shot EPI sequences were acquired. For each sequence and pattern, geometric distortions were quantified in the magnitude images and expressed in millimeters. Phase images were used for extracting the temperature maps on the basis of the temperature-dependent proton resonance frequency shift phenomenon. The means, standard deviations, and signal-to-noise ratios (SNRs) were extracted and contrasted with the geometric distortions of all patterns.ResultsThe geometric distortion analysis and temperature map evaluations showed that more than one pattern could be considered the best-performing transducer. In the sagittal plane, the star (d) (3.46 ± 2.33 mm) and star-ring patterns (f) (2.72 ± 2.8 mm) showed smaller geometric distortions than the currently available seven-segment sheet (c) (5.54 ± 4.21 mm) and were both comparable to the reference scenario (a) (2.77 ± 2.24 mm). Contrasting these results with the temperature maps revealed that (d) performs as well as (a) in SPGR and EPI.ConclusionsWe demonstrated that segmenting the transducer ground plane into a star pattern reduces eddy currents to a level wherein multi-plane EPI for accurate MR thermometry in tcMRgFUS is feasible.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.