The performance and the characteristics of an interferometric double-mirror spectrometer with uncollimated light are studied by measuring spectra of different radiation sources. The stationary interferometer is fabricated without a beam splitter or moving components. The measured interferogram visibilities, which are limited by the size of the source aperture, agree with the theoretical predictions for a slit and a circular source aperture. By background subtraction the effect of detection nonuniformity can be radically reduced to increase the dynamics and the resolving power of the spectrometer. We used a mercury pencil lamp for measurement and found that the dynamic range was ~80 dBm. When isolated spectral lines are measured, the resolving power can be improved by squeezing more than half of a spatial interference cycle onto one pixel. The maximum resolving power reached in measuring the spectra of a diode laser was 1600. The instrument is applicable to a wide range of measurements, such as the recording of temporally variant, wideband radiation sources and the monitoring of laser wavelength.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.