This study presents a respiration sensor that is dependent on a parallel capacitor, including connection lines and electrodes embroidered on textiles. First, characterizations of the respiration capacitor using a silver thread, including a combination of porous Eco-flex simulating air in the lungs due to respiration, were evaluated using an LCR meter. Second, the effects of air gaps on the detection of respiration motions according to the change in electrode distance under pressure were presented. The data values were measured from 1 to 300 kHz using an LCR meter and dielectric test fixture. Third, actual breathing was examined in four patterns: normal breathing, deep breathing, hyperventilation, and apnea. The test was performed after fabricating a clothing-type breathing sensor. Finally, the change in capacitance for actual respiration was determined by wearing a clothing-type respiration sensor based on the data collected. The effectiveness of the respiration sensor was demonstrated by measuring it to discern all waveforms, cycles, and ranges associated with the breathing pattern.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.