Purpose
To compare a biometer using swept-source optical coherence tomography (SS-OCT) with a partial coherence interferometry (PCI)-based biometer in measurements of two ocular biometry parameters, i.e., the axial length and anterior cornea curvature.
Methods
We compared the two biometers SS-OCT (ANTERION, Heidelberg Engineering Inc., Heidelberg, Germany) and PCI (IOL Master, Carl Zeiss Meditec, Jena, Germany) in terms of the axial length (AL) and corneal curvature (K) measurements of 175 eyes. Paired t-tests were used to compare the two biometers. Agreement between the biometers was evaluated using the Bland–Altman method.
Results
The mean age was 36.0 ± 25.6 years (range: 5 to 85 years). The mean axial length was 24.42 ± 0.13 mm for SS-OCT and 24.45 ± 0.14 mm for PCI. The mean corneal curvature was significantly different between the two biometry in flat K (K1) but not in steep K (K2). The limit of agreement was -0.15 to 0.21 in the axial length, -1.18 to 0.83 in K1, and -1.06 to 0.95 in K2. All above ocular biometric measurements between SS-OCT and PCI correlated significantly (Pearson's correlation, p<0.001).
Conclusions
The axial length measured using SS-OCT is useful in clinical practice. It shows a good correlation and agreement with that measured using PCI. However, the axial length and corneal curvature measured using SS-OCT cannot be used interchangeably with that measured using PCI in clinical practice.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.