Purpose Cardiorenal syndrome describes disorders of the heart and the kidneys in which a dysfunction of 1 organ induces a dysfunction in the other. This work describes the design, evaluation, and application of a 4/4‐channel hydrogen‐1/sodium (1H/23Na) RF array tailored for cardiorenal MRI at 7.0 Tesla (T) for a better physiometabolic understanding of cardiorenal syndrome. Methods The dual‐frequency RF array is composed of a planar posterior section and a modestly curved anterior section, each section consisting of 2 loop elements tailored for 23Na MR and 2 loopole‐type elements customized for 1H MR. Numerical electromagnetic field and specific absorption rate simulations were carried out. Transmission field (B1+) uniformity was optimized and benchmarked against electromagnetic field simulations. An in vivo feasibility study was performed. Results The proposed array exhibits sufficient RF characteristics, B1+ homogeneity, and penetration depth to perform 23Na MRI of the heart and kidney at 7.0 T. The mean B1+ field for sodium in the heart is 7.7 ± 0.8 µT/√kW and in the kidney is 6.9 ± 2.3 µT/√kW. The suitability of the RF array for 23Na MRI was demonstrated in healthy subjects (acquisition time for 23Na MRI: 18 min; nominal isotropic spatial resolution: 5 mm [kidney] and 6 mm [heart]). Conclusion This work provides encouragement for further explorations into densely packed multichannel transceiver arrays tailored for 23Na MRI of the heart and kidney. Equipped with this technology, the ability to probe sodium concentration in the heart and kidney in vivo using 23Na MRI stands to make a critical contribution to deciphering the complex interactions between both organs.
Diffusion-weighted imaging (DWI) provides information on tissue microstructure. Single-shot echo planar imaging (EPI) is the most common technique for DWI applications in the brain, but is prone to geometric distortions and signal voids. Rapid acquisition with relaxation enhancement [RARE, also known as fast spin echo (FSE)] imaging presents a valuable alternative to DWI with high anatomical accuracy. This work proposes a multi-shot diffusion-weighted RARE-EPI hybrid pulse sequence, combining the anatomical integrity of RARE with the imaging speed and radiofrequency (RF) power deposition advantage of EPI. The anatomical integrity of RARE-EPI was demonstrated and quantified by center of gravity analysis for both morphological images and diffusion-weighted acquisitions in phantom and in vivo experiments at 3.0 T and 7.0 T. The results indicate that half of the RARE echoes in the echo train can be replaced by EPI echoes whilst maintaining anatomical accuracy. The reduced RF power deposition of RARE-EPI enabled multiband RF pulses facilitating simultaneous multi-slice imaging. This study shows that diffusion-weighted RARE-EPI has the capability to acquire high fidelity, distortion-free images of the eye and the orbit. It is shown that RARE-EPI maintains the immunity to B inhomogeneities reported for RARE imaging. This benefit can be exploited for the assessment of ocular masses and pathological changes of the eye and the orbit.
Shortened dipole antennas based on rectangular dielectric blocks play an important role in ultrahigh field magnetic resonance imaging (UHF-MRI) radio frequency (RF) coil design. However, the generally assumed direct contact with the subject is difficult to maintain in typical in vivo settings. We have previously observed that certain dielectrically shortened dipole antennas can produce a substantially altered transmit field distribution with a very low transmit efficiency when the block and the sample are physically separated. Therefore, the aim of this study was to determine a) why certain designs of dielectrically shortened dipole antennas can produce an inefficient transmit field when the block and the sample are physically separated and b) how this depends on key parameters such as rectangular block geometry, dielectric constant, loading geometry, and RF feeding. In this work, two main types of quasi-transverse dielectric modes were found in different rectangular block geometries and interpreted as TE11δz (MR efficient) and TE1δδy (MR inefficient), and their impact on in vivo MRI experiments involving the human head, calf, and wrist was explored. This study shows, for the first time, why certain antennas preserve their transmit field efficiency despite physical separation from the sample. We conclude that the proposed approach has the potential to provide new insights into dipole antenna design for UHF-MRI.
Three most important eye compartments in the context of sodium physiology were clearly delineated in all of the images: the vitreous humor, the aqueous humor, and the lens. Our results provide encouragement for further clinical studies. The implications for research into eye diseases including ocular melanoma, cataract, and glaucoma are discussed. Magn Reson Med 80:672-684, 2018. © 2018 International Society for Magnetic Resonance in Medicine.
Purpose Potassium ions (K+) play a critical role in cardiac electrophysiology, and changes in their concentration reflect pathophysiological processes related to cardiovascular diseases. Here, we investigated the feasibility of in vivo 39K MRI of the human heart. To achieve this, we developed, evaluated, and applied a 39K/1H RF coil, which is tailored for 39K MRI of human heart at 7.0T. Methods The performance of the 39K/1H RF coil was evaluated by electromagnetic field and specific absorption ratio simulations using 2 (male/female) human voxel models. The RF coil was evaluated at the bench and applied in an in vivo proof‐of‐principle study involving 7 healthy volunteers. The experiments were performed using a 7.0T whole‐body MR system in conjunction with a 3D density‐adapted projection reconstruction imaging technique. Results For in vivo 39K MRI of the human heart, a nominal spatial resolution of 14.5 × 14.5 × 14.5 mm3 within a total scan time of 30 min was achieved. The average SNR within the heart was 9.6 ± 2.4. Conclusion This work validates the design of a 39K/1H RF coil for cardiac MR at 7.0T and demonstrates for the first time in vivo the feasibility of 39K MRI of the human heart.
Purpose: The goal of the study is to develop 31P spectroscopic MR fingerprinting (MRF) at 7T to measure T1 and T2 relaxation times simultaneously and to compare time efficiency and test-retest reproducibility of MRF with conventional inversion recovery and multi-TE methods. Methods: A 31P MRF scheme was designed based on a balanced steady-state free precession type sequence. Dictionary was generated using the Bloch equations. B0 map was acquired experimentally and incorporated into the dictionary. Simulations were performed to evaluate estimation bias. 7 phantoms with different T1 and T2 relaxation times were prepared for MRF validation. 7 volunteers were scanned twice using both MRF and the conventional methods to evaluate the reproducibility. Results: In phantom measurements, T1 and T2 values between MRF and conventional methods demonstrated a good agreement with Pearson's correlation coefficients of 0.99 and 0.97, respectively. In in vivo experiments, estimated T1 by MRF were in good agreement with those measured by the inversion recovery and in literatures. On the other hand, estimated T2 values by MRF were shorter than those measured by the multi-TE method. 31P MRF method can reduce the acquisition time by 15 min providing less than 10% of mean CV for T1 estimations and less than 20% of mean CV for T2 estimations of metabolites. Conclusion: Our results shows the feasibility of simultaneous T1 and T2 measurements of 31P metabolites in human brain using the MRF technique at 7T. High reproducibility can be achieved especially for T1 measurement with 40% time reduction over conventional methods.
Purpose In this study, we aimed to measure the apparent diffusion coefficients (ADCs) of major phosphorous metabolites in the human calf muscle at 7 T with a diffusion-weighted (DW)-STEAM sequence. Methods A DW-STEAM sequence with bipolar gradients was implemented at 7 T, and DW MR spectra were acquired in three orthogonal directions in the human calf muscle of six healthy volunteers (TE/TM/TR = 15 ms/750 ms/5 s) at three b-values (0, 800, and 1200 s/mm2). Frequency and phase alignments were applied prior to spectral averaging. Averaged DW MR spectra were analyzed with LCModel, and ADCs of 31P metabolites were estimated. Results Four metabolites (phosphocreatine (PCr), adenosine triphosphate (ATP), inorganic phosphate (Pi) and glycerol phosphorylcholine (GPC)) were quantified at all b-values with mean CRLBs below 10%. The ADC values of PCr, ATP, Pi, and GPC were (0.24 ± 0.02, 0.15 ± 0.04, 0.43 ± 0.14, 0.40 ± 0.09) × 10–3 mm2/s, respectively. Conclusion The ADCs of four 31P metabolites were successfully measured in the human calf muscle at 7 T, among which those of ATP, Pi and GPC were reported for the first time in humans. This study paves the way to investigate 31P metabolite diffusion properties in health and disease on the clinical MR scanner.
Objective To determine whether a multi-feed, loop-dipole combined approach can be used to improve performance of rectangular dielectric resonator antenna (DRA) arrays human brain for MRI at 7 T. Materials and methods Electromagnetic field simulations in a spherical phantom and human voxel model “Duke” were conducted for different rectangular DRA geometries and dielectric constants εr. Three types of RF feed were investigated: loop-only, dipole-only and loop-dipole. Additionally, multi-channel array configurations up to 24-channels were simulated. Results The loop-only coupling scheme provided the highest B1+ and SAR efficiency, while the loop-dipole showed the highest SNR in the center of a spherical phantom for both single- and multi-channel configurations. For Duke, 16-channel arrays outperformed an 8-channel bow-tie array with greater B1+ efficiency (1.48- to 1.54-fold), SAR efficiency (1.03- to 1.23-fold) and SNR (1.63- to 1.78). The multi-feed, loop-dipole combined approach enabled the number of channels increase to 24 with 3 channels per block. Discussion This work provides novel insights into the rectangular DRA design for high field MRI and shows that the loop-only feed should be used instead of the dipole-only in transmit mode to achieve the highest B1+ and SAR efficiency, while the loop-dipole should be the best suited in receive mode to obtain the highest SNR in spherical samples of similar size and electrical properties as the human head.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.