A monolithic methacrylate packed 96-tips device was used for the extraction of the busulfan and cyclophosphamide in whole blood samples. Using a packed 96-tips set, a 96-well plate could be handled in about 2 min. The key aspect of the monolithic phase is that monolithic material can offer both good extraction capacity and low-back-pressure properties. The validation of the methodology showed that the accuracy values of quality-control samples were between 99 and 113% for busulfan, and between 103 and 110% for cyclophosphamide. The inter-day precision ranged from 7.0 to 12% for busulfan and from 13 to 16% for cyclophosphamide. The standard calibration curves were obtained within the concentration range 5-2000 nm for busulfan and from 10 to 5000 nm for cyclophosphamide in blood samples. The coefficients of determination were ≥0.99.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.