Chromatographic resolution and analysis accuracy can be seriously affected by broadening of very sharp bands associated with both high-performance size-exclusion chromatography and other liquid chromatography methods using less than 5 micron particles. The effect on column plate count by extra-column band broadening from injection systems, detectors, connectors and "guard" columns is discussed and exemplified with experimental data. The special cases of tailing and doublet peaks for a single species are investigated. These effects are eliminated by a special sample injection technique. Computer simulation studies demonstrated that serious errors can result from calculating column plate count with non-Gaussian peaks. The maximum allowable peak tailing is suggested and a new method for quantitating peak skew is proposed.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.