Powered resistive and resistive-superconductive hybrid magnets can reach fields higher than superconducting NMR magnets but lack the field homogeneity and temporal stability needed for high resolution NMR. Due to field fluctuations in powered magnets, commercially available mapping systems fail to produce maps of these magnets with sufficient reproducibility, thus hampering attempts to improve homogeneity of the field they generate. Starting with a commercial mapper, we built a mapping system which uses a two-channel (measurement + reference) mapper probe. We used this system to map and then to shim two magnets of Florida Bitter type at the National High Magnetic Field Laboratory in Tallahassee, FL. With a combination of passive (ferromagnetic) and active shims we achieved 2.3 ppm homogeneity in 1 cm diameter spherical volume (dsv) at 25.0 T in the Keck resistive magnet, and 0.9 ppm homogeneity in 1 cm dsv at 23.5, 28.2, and 35.2 T in the series-connected resistive-superconductive hybrid (SCH) magnet.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.