2017
DOI: 10.1016/j.nima.2017.09.002
|View full text |Cite
|
Sign up to set email alerts
|

Sectioned calorimeter for quick diagnostic of the electron beam energy distribution

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1

Citation Types

0
5
0

Year Published

2018
2018
2022
2022

Publication Types

Select...
8
1

Relationship

0
9

Authors

Journals

citations
Cited by 10 publications
(5 citation statements)
references
References 16 publications
0
5
0
Order By: Relevance
“…Paper describe design of a calorimeter with infrared registration of temperature field in sectioned collector, the design is tested with an accelerator described in paper [8]. The modified device design is based on construction from [10]. The device setup is depicted on Fig.…”
Section: Sectioned Calorimeter With Current Detector For Studying The...mentioning
confidence: 99%
See 1 more Smart Citation
“…Paper describe design of a calorimeter with infrared registration of temperature field in sectioned collector, the design is tested with an accelerator described in paper [8]. The modified device design is based on construction from [10]. The device setup is depicted on Fig.…”
Section: Sectioned Calorimeter With Current Detector For Studying The...mentioning
confidence: 99%
“…Convective heat transfer was not taken into account for experiments under vacuum, because the air pressure in vacuum chamber is rather low (around 10 -2 Pa). For experiments under atmosphere, it is shown that an estimation of beam energy distribution may be performed with an accuracy around 10% in interval up to 10 s since the calorimeter irradiation end [10].…”
Section: Sectioned Calorimeter With Current Detector For Studying The...mentioning
confidence: 99%
“…However, to provide a spatial resolution better than 1 mm for a beam area that exceeds 20 cm 2 , the calorimeter design becomes complex and the measurement process takes a long time. A sectioned calorimeter designed for registering an electron beam was described in [53]. The collector of the calorimeter, with a diameter of 100 mm, has 61 sections, whose temperature was determined using a thermal imager.…”
Section: The Diagnostics Of Electron Beamsmentioning
confidence: 99%
“…The energy of a pulsed electron beam taken out into the atmosphere was measured with a help of regular and sectioned calorimeter [22]. The regular calorimeter consists of a copper disk 2 mm x Ø90 mm and a temperature sensor.…”
Section: Fig 2 Scheme Of the Experiments On Measuring The Beam Energymentioning
confidence: 99%