Optimization and Long-Term Stability of Micro Flow Sensors for Smart VP Shunts

Abstract: This paper reports on a systematic study of the flow sensitivity and resolution of micro-flow sensors intended for use in implantable ventricular-peritoneal shunts. The flow sensors utilize ferromagnetic flaps (transducers) whose deflection is detected by ultra-sensitive MTJ sensors (20mV/V/Oe). A working range of 0-40 ml/h was demonstrated at a maximum uncertainty of 4% RMS and a resolution of 0.4 ml/h. Earlier studies on this sensor unveiled significant low-frequency noise (drift) limiting the sensitivity to… Show more