2016
DOI: 10.1118/1.4947293
|View full text |Cite
|
Sign up to set email alerts
|

How doesc-viewimage quality compare with conventional 2D FFDM?

Abstract: Their analysis demonstrates many instances where the c-view image quality differs from FFDM. Compared to FFDM, c-view offers a better depiction of objects of certain size and contrast, but provides poorer overall resolution and noise properties. Based on these findings, the utilization of c-view images in the clinical setting requires careful consideration, especially if considering the discontinuation of FFDM imaging. Not explicitly explored in this study is how the combination of DBT + c-view performs relati… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

6
51
3
2

Year Published

2017
2017
2023
2023

Publication Types

Select...
6

Relationship

0
6

Authors

Journals

citations
Cited by 67 publications
(62 citation statements)
references
References 14 publications
6
51
3
2
Order By: Relevance
“…We found that this also holds true for INSIGHT3D without significant differences with respect to SM or the standard stack reconstruction. Previous studies about SM described a worse depiction of the morphology of microcalcifications compared to FFDM [7,12]. This comparison, however, was not part of this study and describing the morphology with either of these reconstruction techniques might not be comparable with FFDM.…”
Section: Discussioncontrasting
confidence: 43%
See 2 more Smart Citations
“…We found that this also holds true for INSIGHT3D without significant differences with respect to SM or the standard stack reconstruction. Previous studies about SM described a worse depiction of the morphology of microcalcifications compared to FFDM [7,12]. This comparison, however, was not part of this study and describing the morphology with either of these reconstruction techniques might not be comparable with FFDM.…”
Section: Discussioncontrasting
confidence: 43%
“…While the advantages of SM, such as better and faster comparability to previous examinations and contralateral imaging [6], are preserved with INSIGHT3D, its superior depiction of microcalcifications shows added value to SM. In this context, however, it should be remembered that it is not yet clear whether SM can generally substitute FFDM for the depiction of microcalcifications [7,11,12].…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…The detector was used in full resolution mode for all measurements. In this study, the wired test object resting on the detector was used as the reference (wire height 10 mm) MTF 10 . At 5 cm above the detector, the MTFs of the central projection images in both the chest wall, and scanning direction had an approximate 23% reduction in spatial resolution, showing no noticeable reduction in resolution of one direction over the other.…”
Section: E1 Mtf Of Dbt Projection Imagesmentioning
confidence: 99%
“…Characteristics such as distribution, morphology, and size are important for determining malignancy. As a result, FDA approved systems are required to be used in conjunction with 2D mammography . Recently, the FDA has approved synthetic 2D images together with DBT as an alternative solution to patients concerned with radiation exposure.…”
Section: Introductionmentioning