2012
DOI: 10.1007/s00167-012-2208-y
|View full text |Cite
|
Sign up to set email alerts
|

Fixed- versus mobile-bearing unicondylar knee arthroplasty: are failure modes different?

Abstract: The available evidence has not confirmed the advantage of mobile-bearing UKAs over fixed-bearing UKAs but pointed out specific modes of failure.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

4
26
0

Year Published

2015
2015
2024
2024

Publication Types

Select...
5
4

Relationship

0
9

Authors

Journals

citations
Cited by 35 publications
(30 citation statements)
references
References 71 publications
4
26
0
Order By: Relevance
“…This design should reduce the contact stress at the articulating interface and result in a low bearing wear rate (0.01-0.04 mm/y) 8,45 than FB UKA (0.15 mm/y). 37,44 Compared with previous meta-analyses, 40,41 we included more updated and new clinical research studies (published up to 2017) and a larger sample size ($50% more patients were included), allowing a more precise comparison of MB UKA and FB UKA. The most significant result of our review was that the FB UKA group showed higher knee scores and superior ROM than the MB UKA group.…”
Section: Discussionmentioning
confidence: 99%
“…This design should reduce the contact stress at the articulating interface and result in a low bearing wear rate (0.01-0.04 mm/y) 8,45 than FB UKA (0.15 mm/y). 37,44 Compared with previous meta-analyses, 40,41 we included more updated and new clinical research studies (published up to 2017) and a larger sample size ($50% more patients were included), allowing a more precise comparison of MB UKA and FB UKA. The most significant result of our review was that the FB UKA group showed higher knee scores and superior ROM than the MB UKA group.…”
Section: Discussionmentioning
confidence: 99%
“…A large number of studies report wearing at late period for UKA with fixed-bearing insert and insert luxation at early period for UKA with mobile-bearing insert (17)(18)(19). Patellofemoral joint is not mentioned sufficiently despite a change in indications.…”
Section: Discussionmentioning
confidence: 99%
“…Several systematic reviews on this topic have been published recently. Cheng et al 10 found equivalent range of motion, limb alignment, patient-reported outcomes, incidence of aseptic loosening, and reoperation rates between the two bearing designs. However, the time to reoperation and failure mode were different.…”
Section: Fixed Vs Mobile Bearing Designmentioning
confidence: 97%