External Ventricular Catheters: Is It Appropriate to Use an Open/Monitor Position to Adequately Trend Intracranial Pressure in a Neuroscience Critical Care Environment?

Abstract: Intracranial pressure (ICP) monitoring can be an important assessment tool in critically and acutely ill patients. An external ventricular drain offers a comprehensive way to monitor ICP and drain cerebrospinal fluid. The Monro-Kellie hypothesis, Pascal's principle, and fluid dynamics were used to formulate an assumption that an open/monitor position on the stopcock is an adequate trending measure for ICP monitoring while concurrently draining cerebrospinal fluid. Data were collected from 50 patients and total… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

0
5
0

Year Published

2017
2017
2021
2021

Publication Types

Select...
3

Relationship

0
3

Authors

Journals

citations
Cited by 3 publications
(5 citation statements)
references
References 0 publications
0
5
0
Order By: Relevance
“…The nursing staff must monitor signs indicating infection, bleeding, document aspects of the drainage system, in addition to providing emotional support to patients' family (29) . Moreover, sometimes nurses perform CSF drainage as directed, to decrease CSF when elevated (30) .…”
Section: Discussionmentioning
confidence: 99%
“…The nursing staff must monitor signs indicating infection, bleeding, document aspects of the drainage system, in addition to providing emotional support to patients' family (29) . Moreover, sometimes nurses perform CSF drainage as directed, to decrease CSF when elevated (30) .…”
Section: Discussionmentioning
confidence: 99%
“…One clinical study involving 50 patients being treated with EVD investigated the accuracy of pressure measurements made from the drainage line with concurrent flow and compared them to the standard of care: Traditional fluid-coupled measurements made by stopping flow [ 6 ]. These authors found that the error due to flow was on average 2.2 cmH 2 O (1.6 mmHg) and was less than 4 cmH 2 O (3 mmHg) 97% of the time, concluding that it was negligible.…”
Section: Discussionmentioning
confidence: 99%
“…1 1, 2 3, 4 5, 1, 6 1 Simultaneous drainage and ICP measurement has not been adopted clinically given concern for sources of error in these pressure measurements, however, the error has not been well-characterized. Proposed benefits of continuous monitoring include monitoring cerebrovascular autoregulation by pressure reactivity index without the need for an additional probe and rapid detection of CSF flow blockages and other emergencies that may place the patient at risk [6,7].…”
Section: Introductionmentioning
confidence: 99%
See 1 more Smart Citation
“…157,[165][166][167] This concern is relevant to transducers that are mounted externally and transducers placed within the tip of the drainage catheter. 166 Although a value for ICP can be trended while the device is simultaneously draining, 168 the EVD/LD should be closed for accurate measurement at least hourly. 166,168 In some situations such as an anesthetized patient with elevated ICP, it would be indicated to obtain accurate pressure measurements more frequently than hourly.…”
Section: Inaccurate Pressure Measurements With Simultaneous Drainagementioning
confidence: 99%