Intracranial pressure (ICP) monitoring can be an important assessment tool in critically and acutely ill patients. An external ventricular drain offers a comprehensive way to monitor ICP and drain cerebrospinal fluid. The Monro-Kellie hypothesis, Pascal's principle, and fluid dynamics were used to formulate an assumption that an open/monitor position on the stopcock is an adequate trending measure for ICP monitoring while concurrently draining cerebrospinal fluid. Data were collected from 50 patients and totaled 1053 separate number sets. The open/monitor position was compared with the clamped position every hour. An order for "open to drain" was needed for appropriate measurement and nursing care. Results showed the absolute average differences between open/monitor and clamped positions at 1.6268 mm Hg. This finding suggests that it is appropriate to use an open/monitor position via an external ventricular drain for adequate trending of patients' ICP.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.