2011
DOI: 10.1118/1.3601018
|View full text |Cite
|
Sign up to set email alerts
|

Dosimetric and mechanical characteristics of a commercial dynamic μMLC used in SRS

Abstract: (a) The add-on microMLC system in combination with our LINAC has been commissioned to be used for clinical purposes and showed good agreement with published results for different ,MLC types. (b) This work has lead to the recommendation that leaves should be recalibrated after ten static beams or after each dynamic arc.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1

Citation Types

0
1
0

Year Published

2014
2014
2017
2017

Publication Types

Select...
4

Relationship

0
4

Authors

Journals

citations
Cited by 4 publications
(1 citation statement)
references
References 14 publications
(18 reference statements)
0
1
0
Order By: Relevance
“…Galal et al tested the dosimetric and mechanical characteristics of a 3-mm micro-MLC and concluded that it was suitable for use in radiosurgery [31]. Many investigators have studied the effects of MLC leaf width in various planning techniques (3D-CRT and IMRT).…”
Section: Discussionmentioning
confidence: 99%
“…Galal et al tested the dosimetric and mechanical characteristics of a 3-mm micro-MLC and concluded that it was suitable for use in radiosurgery [31]. Many investigators have studied the effects of MLC leaf width in various planning techniques (3D-CRT and IMRT).…”
Section: Discussionmentioning
confidence: 99%