Development of a protocol for small beam bi-dimensional dose distribution measurements with radiochromic films

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
7
0

Year Published

2018
2018
2024
2024

Publication Types

Select...
5

Relationship

0
5

Authors

Journals

citations
Cited by 5 publications
(7 citation statements)
references
References 29 publications
0
7
0
Order By: Relevance
“…Moreover, they allow 2D dose measurements and they can easily be housed in anthropomorphic phantoms. Measurements were performed according to a rigorous protocol (cutting, calibration, readout) developed in our laboratory 11 . In particular, it is based on a pixel‐to‐pixel background subtraction method with the use of the red channel only in order to overcome the limits of the multichannel correction method at low doses 15 .…”
Section: Methodsmentioning
confidence: 99%
See 2 more Smart Citations
“…Moreover, they allow 2D dose measurements and they can easily be housed in anthropomorphic phantoms. Measurements were performed according to a rigorous protocol (cutting, calibration, readout) developed in our laboratory 11 . In particular, it is based on a pixel‐to‐pixel background subtraction method with the use of the red channel only in order to overcome the limits of the multichannel correction method at low doses 15 .…”
Section: Methodsmentioning
confidence: 99%
“…In particular, it is based on a pixel‐to‐pixel background subtraction method with the use of the red channel only in order to overcome the limits of the multichannel correction method at low doses 15 . Measurements uncertainty was assessed as described in 11 . This protocol leads to dose measurements with a standard deviation of 2.9% (1 sigma) in the 0.5–4.0 Gy dose range (reaching 4.5% (1 sigma) for doses below 0.5 Gy).…”
Section: Methodsmentioning
confidence: 99%
See 1 more Smart Citation
“…The PENELOPE program (version 2014 and the main program penEasy version 2017) (Salvat 2015) was used to precisely model a 1000 MU min −1 Cyberknife VSI equipped with fixed cylindrical collimators. In this study, the detailed quadric geometry of the Cyberknife head modeled in a previous work (Moignier et al 2014a(Moignier et al , 2016 was completed. This geometry, based on manufacturer drawings and materials, includes the head components shaping the beam line and was previously validated in comparison with films and various detectors measurements for in-field dose evaluation (up to approximatively 6 cm from beam axis).…”
Section: Monte Carlo Simulation Of the Cyberknife For Out-of-field Do...mentioning
confidence: 99%
“…Radiochromic films have been chosen for out-of-field measurements as their response shows little dependence with energy (Massillon-JL et al 2012, Borca et al 2013). A rigorous protocol (Moignier et al 2016), developed in our laboratory, was used for the films' calibration and reading. This protocol enables EBT3 film dose measurements with a standard deviation of 2.9% (1 sigma) in the range from 0.5 to 4.0 Gy (reaching 4.5% (1 sigma) for 0.5 Gy).…”
Section: Radiochromic Films and Detectors Measurementsmentioning
confidence: 99%