2004
DOI: 10.1258/002221504323219518
|View full text |Cite
|
Sign up to set email alerts
|

Complications following adult cochlear implantation: experience in Manchester

Abstract: Cochlear implantation is regarded as a safe and effective treatment for the profoundly deaf. However, a proportion of patients suffer complications after implant surgery. This paper examines the complications encountered in 240 adult cochlear implant operations performed in Manchester between June 1988 and June 2002. Minor complications were defined as those that either settled spontaneously or with conservative management. The total number of minor complications was 61 (25.4 per cent of cases). Non-auditory s… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

5
52
0
7

Year Published

2005
2005
2016
2016

Publication Types

Select...
7
1

Relationship

0
8

Authors

Journals

citations
Cited by 90 publications
(64 citation statements)
references
References 12 publications
5
52
0
7
Order By: Relevance
“…A similar study was carried out in Izmir, Turkey; in a pediatric sample with 227 implantees, the incidence of major and minor complications were 12.33% and 6.6%, respectively 8 . Relatively high incidences have been observed in Manchester from June of 1988 to June of 2002 in a sample of 240 adult implantees, in whom they observed 6.25% and 25.4% of major and minor complications, respectively 9 .…”
Section: Introductionmentioning
confidence: 98%
“…A similar study was carried out in Izmir, Turkey; in a pediatric sample with 227 implantees, the incidence of major and minor complications were 12.33% and 6.6%, respectively 8 . Relatively high incidences have been observed in Manchester from June of 1988 to June of 2002 in a sample of 240 adult implantees, in whom they observed 6.25% and 25.4% of major and minor complications, respectively 9 .…”
Section: Introductionmentioning
confidence: 98%
“…The overall incidence of complications is reported as between 7-37 per cent in the literature. 12 In our series of patients the overall rate of complications related to surgery itself was 5.77 per cent. There was one patient who had transient wound swelling post-operatively.…”
Section: Discussionmentioning
confidence: 55%
“…30 Elektrotun yanlış yerleştirilmesi ve hasarı varsa;bu, operasyon sırasında ya da sonrasında çe-kilen röntgen filmleriyle saptanabilir ve elektrotun yeniden yerleştirilmesini gerektirir. 30,41,42 Flep nekrozu varsa; implant çıkarılmalı ve cerrah tarafından ortalama bir yıl sonra implant tekrarlanmalı; vasküler desteği bozmayacak, venöz ve lenfatik drenajı engellemeyecek yeni bir flep tasarlanmalıdır. 30 Ameliyattan sonra hastanede yatış genellikle 24 saat 10 gün arasında değişmektedir.…”
Section: Ameli̇yat Sonrasi Hasta Bakimiunclassified
“…6,22,34,[43][44][45][46][47] Kohlear implant cerrahisi komplikasyonları; ameliyat sırası ve ameliyat sonrası dönemlerde görülme durumlarına göre gruplandırılarak aşağıda açıklan-mıştır. 6,22,30,44,47,48 A Am me el li iy ya at t s sı ır ra as sı ın nd da a e en n s sı ık k g gö ör rü ül le en n k ko om mp pl li ik ka as s--y yo on nl la ar r; ; ■ Hemoraji, 6,30 ■ Fasiyal sinir ve korda timpani zedelenmesi, 6,30,44 ■ Perilenf sızıntısı, 6,30 ■ Elektrotun yanlış yerleştirilmesi, 6,22,41,46,47 ■ Elektrot hasarı. 6,22,30,47 A Am me el li iy ya at t s so on nr ra as sı ı d dö ön ne em md de e e en n ç ço ok k k ka ar rş şı ıl la aş şı ıl la an n k ko om mp pl li ik ka as sy yo on nl la ar r; ; ■ Hematom, 6,23,30,42,45,46,48 ■ Seroma, 6,30,43 ■ Aerosel, 6,30 ■ Beyin; omurilik sıvısı sızıntısı, 23,42,44,…”
Section: Kohlear İmplant Cerrahi̇si̇ni̇n Kompli̇kasyonlariunclassified
See 1 more Smart Citation