2016
DOI: 10.1016/j.ijrobp.2016.03.033
|View full text |Cite
|
Sign up to set email alerts
|

Comparison of Dose Distributions With TG-43 and Collapsed Cone Convolution Algorithms Applied to Accelerated Partial Breast Irradiation Patient Plans

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

2
4
0

Year Published

2017
2017
2021
2021

Publication Types

Select...
8

Relationship

0
8

Authors

Journals

citations
Cited by 10 publications
(6 citation statements)
references
References 19 publications
2
4
0
Order By: Relevance
“…Our calculations found TG-186-calculated DVH-parameters to be in general lower than TG-43U1calculated parameters. This is in good agreement with the results of other groups who reported an overestimation of TG-43U1 in comparison to TG-186 for soft tissue in other tumour entities [4,6,8,10] [8]. For cervix cancer patients, Jacob et al compared TG-43 to D m,m and reported changes in target coverage parameter D 90 of 2.6%, with similar values for D 95 and D 100 .…”
Section: Discussionsupporting
confidence: 90%
See 1 more Smart Citation
“…Our calculations found TG-186-calculated DVH-parameters to be in general lower than TG-43U1calculated parameters. This is in good agreement with the results of other groups who reported an overestimation of TG-43U1 in comparison to TG-186 for soft tissue in other tumour entities [4,6,8,10] [8]. For cervix cancer patients, Jacob et al compared TG-43 to D m,m and reported changes in target coverage parameter D 90 of 2.6%, with similar values for D 95 and D 100 .…”
Section: Discussionsupporting
confidence: 90%
“…Other groups also suggested a correlation between deviations from TG-43U1 and photon travel distance path. Thrower et al and Zourari et al found that deviations increased with increasing distance from dwell positions [4,5] for breast cases, although in that case this was mostly due to lung heterogeneities.…”
Section: Liver Dosementioning
confidence: 92%
“…using AcurosBV (calculating D w,m ) for ten patients treated with flexible plastic catheters [41], and by Thrower et al . using ACE (calculating D m,m ) for 50 patients treated with a SAVI or Contura device [43]. These studies found dose differences for the CTV and OARs relative to TG-43 to be consistent with those of earlier investigations, insofar as direct comparisons were possible.…”
Section: Experience Using Commercial Model-based Dose Calculation Algsupporting
confidence: 66%
“…AAPM has published Task Group 186, which guides beginners regarding MBDCAs for BT dose calculations and ensures uniformity in practice. Previous papers have already demonstrated the efficacy of MBDCAs for phantoms and for several treatment sites, such as breast tissue [ 15 19 ]. However, to date, there has been minimal evaluation of the impact of MBDCAs on the treatment of cervical cancer, particularly for the Advanced Collapsed cone Engine (ACE) method [ 16 , 20 ].…”
Section: Introductionmentioning
confidence: 99%