Comparative performance analysis of Urised 3 and DIRUI FUS-200 automated urine sediment analyzers and manual microscopic method

Abstract: Objective: Microscopic examination of urine sediment is necessary for evaluation of renal and urinary tract diseases. In this study, we evaluated and compared analytic and diagnostic performances of DIRUI FUS-200 and a new image-based automated urine sediment analyzer Urised 3. Method: A total of 440 urine samples, submitted to our laboratory, were evaluated by two automated urine sediment analyzers and a standardized manual microscopy. Precision, linearity and method comparison studies were performed accordin… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4
1

Citation Types

0
8
0

Year Published

2020
2020
2023
2023

Publication Types

Select...
5

Relationship

0
5

Authors

Journals

citations
Cited by 5 publications
(8 citation statements)
references
References 22 publications
0
8
0
Order By: Relevance
“…In addition, Bland-Altman agreement plot graphics demonstrated that automated microscopy units of the two devices showed acceptable performance for WBC (-29.3 ± 1.96 SD), RBC (43.3 ± 1.96 SD) and EPI (-4.0 ± 1.96 SD) cell counts [18] . Yalçınkaya et al [3] compared FUS200 and Urised3 urine analyzers where Deming regression analysis yielded a correlation coefficient of 0.961 and 0.961 for WBC and RBC counts, respectively. When they investigated the consistency between two urine analyzers for negative-positive test results, they found the kappa value to be 0.79 (good consistency) for WBC and 0.42 (moderate consistency) for RBC [3] .…”
Section: Discussionmentioning
confidence: 99%
See 2 more Smart Citations
“…In addition, Bland-Altman agreement plot graphics demonstrated that automated microscopy units of the two devices showed acceptable performance for WBC (-29.3 ± 1.96 SD), RBC (43.3 ± 1.96 SD) and EPI (-4.0 ± 1.96 SD) cell counts [18] . Yalçınkaya et al [3] compared FUS200 and Urised3 urine analyzers where Deming regression analysis yielded a correlation coefficient of 0.961 and 0.961 for WBC and RBC counts, respectively. When they investigated the consistency between two urine analyzers for negative-positive test results, they found the kappa value to be 0.79 (good consistency) for WBC and 0.42 (moderate consistency) for RBC [3] .…”
Section: Discussionmentioning
confidence: 99%
“…Yalçınkaya et al [3] compared FUS200 and Urised3 urine analyzers where Deming regression analysis yielded a correlation coefficient of 0.961 and 0.961 for WBC and RBC counts, respectively. When they investigated the consistency between two urine analyzers for negative-positive test results, they found the kappa value to be 0.79 (good consistency) for WBC and 0.42 (moderate consistency) for RBC [3] . In the present study, Bland-Altman analysis revealed lower mean of differences between manual microscopy and UriSed2 device for WBC, RBC, EPI (-9.8, -13.5 and 7.7, respectively) suggesting that manual microscopy and UriSed2 device are concordant.…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…Thus, automated urine autoanalyzers have been used in clinical laboratories with higher patient potential. To deal with the disadvantages of manual microscopy, automated urine analyzers using several analytical techniques have been developed [7]. Currently, these devices are frequently used in medical laboratories to minimize the work force lost during manual microscopic examination.…”
Section: Discussionmentioning
confidence: 99%
“…In a study performed by Yalçınkaya et al with 440 urine samples based on the comparison of FUS200 and Urised3 microscopic urine analyzers with the Deming regres- sion analyzer, the correlation coefficients for WBC and RBC measurements performed by the two devices were 0.961 and 0.961, respectively. When they examined the concordance between the two microscopic analyzers for negative-positive test results, they found the Kappa value to be 0.79 (good agreement) for WBC and 0.42 (moderate agreement) for RBC [14]. In a study performed by İnigo et al with 1934 urine samples to compare the results obtained by SediMax and Sysmex UF1000i analyzers, they performed the Spearman correlation analysis between the devices and found a very strong correlation for WBC (r=0.928) and a strong correlation for RBC (r=0.631) and Squamous epithelium cell counts (r=0.631) between the devices [15].…”
Section: Discussionmentioning
confidence: 99%