A Comparative study on the performances of 77 Elektronika Urised 2-LabUmat 2 and Dirui FUS 200-H 800 urine analyzers

Abstract: A comparative study on the performances of 77 elektronika urised 2-LabUmat2 and Dirui FUS200 -H800 urine analyzersObjectives: This study aims to compare the analytical performance of the 77Elektronika brand LabUmat2-Urised2 model integrated urine analyzer and Dirui brand FUS200-H800 model integrated urine analyzer. Methods: Urine samples of 139 patients randomly selected from male and female patients of all age groups who were admitted to Ufuk University Faculty of Medicine, Dr. Ridvan Ege Hospital between 24.… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3

Citation Types

0
3
0

Year Published

2022
2022
2022
2022

Publication Types

Select...
1

Relationship

0
1

Authors

Journals

citations
Cited by 1 publication
(3 citation statements)
references
References 14 publications
(23 reference statements)
0
3
0
Order By: Relevance
“…In the same study, epithelial cell count demonstrated consistency by 86.6% between UF-1000i and iQ200, 85.1% between UF-1000i and UriSed, 87.6% between iQ200 and UriSed, and 90.2% between UF-1000i and iQ200 [16] . In a study, Ercin [18] compared Urised2 and FUS200 microscopic analyzers and found WBC, RBC and EPI cell counts to be within the same range and reported excellent consistency for these three parameters (Gamma value is 0.916, 0.770 and 0.961, respectively; p <0.001). Jintasuthanont et al [10] compared the results of UriSed analyzer with that of manual microscopy; they found the gamma value to be 0.837 for WBC, 0.918 for RBC, and 0.939 for squamous epithelial cell count, and they detected high correlation between the results obtained by full-automated analyzer and the results obtained by manual microscopy.…”
Section: Discussionmentioning
confidence: 93%
See 2 more Smart Citations
“…In the same study, epithelial cell count demonstrated consistency by 86.6% between UF-1000i and iQ200, 85.1% between UF-1000i and UriSed, 87.6% between iQ200 and UriSed, and 90.2% between UF-1000i and iQ200 [16] . In a study, Ercin [18] compared Urised2 and FUS200 microscopic analyzers and found WBC, RBC and EPI cell counts to be within the same range and reported excellent consistency for these three parameters (Gamma value is 0.916, 0.770 and 0.961, respectively; p <0.001). Jintasuthanont et al [10] compared the results of UriSed analyzer with that of manual microscopy; they found the gamma value to be 0.837 for WBC, 0.918 for RBC, and 0.939 for squamous epithelial cell count, and they detected high correlation between the results obtained by full-automated analyzer and the results obtained by manual microscopy.…”
Section: Discussionmentioning
confidence: 93%
“…Ercin [18] reported high correlation between Urised2-LabUmat2 and FUS200-H800 devices for WBC count, but moderate correlation for RBC count. Moreover, according to the Passing-Bablok regression analysis in this study, they reported that there is no consistency between these two methods because of the presence of remarkable deviation from linearity for WBC and RBC counts (p <0.05 and p=0.01, respectively) but that they are concordance for EPI cell count (p=0.65).…”