Experimental
ApparatusThe violet-LED (light-emitting diode) light source (∼2 mW), CE set-up and data-acquisition system used were similar to that described previously, 10 and are abbreviated herein. A large-volume sample injection (> 5 µL) with an extremely high theoretical plate number (N > 10 7 ) was achieved when the sweeping-MEKC mode and a coupled-capillary (100 -50 µm i.d.) were simultaneously used in a capillary electrophoresis (CE) separation. A low-cost and compact violet-LED (∼2 mW) was used as the fluorescence excitation source. As a result, the theoretical plate numbers of the detected peaks (two model compounds: naphthalene-2,3-dicarboxaldehyde derivatized-dopamine and -norepinephrine) were 1.0 × 10 7 and 7.4 × 10 6 , respectively. The limits of detection (at S/N = 3) of these were determined to be 2.8 × 10 -10 M (92 ppt) and 2.3 × 10 -10 M (83 ppt), respectively.