An Outbreak of Pyrogenic Reactions in Chronic Hemodialysis Patients Associated with Hemodialyzer Reuse

Abstract: In February 1992, 22 patients undergoing chronic hemodialysis at an outpatient dialysis center experienced pyrogenic reactions (PR). The PR rate was significantly greater (p < 0.001) during the epidemic (February 3-5) than the pre-epidemic period (November 1, 1992-February 1, 1992). All patients with PR used dialyzers that had been manually reprocessed either on February 1 or 3. These dialyzers contained up to 120.8 EU/ml of endotoxin in the blood compartment. The only dialyzer reprocessed before February 1 th… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
14
0
2

Year Published

1997
1997
2023
2023

Publication Types

Select...
4
3
1

Relationship

3
5

Authors

Journals

citations
Cited by 27 publications
(16 citation statements)
references
References 7 publications
0
14
0
2
Order By: Relevance
“…The procedure remains potentially hazardous to patients, and numerous clusters of gram-negative bacteremia and pyrogenic reactions (PR) have been reported. Most such clusters have been traced to bacterial contamination of water and/or dialysate [3,4], errors in dialyzer reprocessing [5][6][7], and improper setup procedures [8][9][10]. In this paper, we describe an outbreak of Enterobacter cloacae bloodstream infections and PR at Hôpital Charles-LeMoyne, a 590-bed university-affiliated teaching hospital in the Montreal area, that was first detected in August 1995 when five episodes of bacteremia occurred among the dialysis unit's 60 chronic hemodialysis patients.…”
Section: Introductionmentioning
confidence: 99%
“…The procedure remains potentially hazardous to patients, and numerous clusters of gram-negative bacteremia and pyrogenic reactions (PR) have been reported. Most such clusters have been traced to bacterial contamination of water and/or dialysate [3,4], errors in dialyzer reprocessing [5][6][7], and improper setup procedures [8][9][10]. In this paper, we describe an outbreak of Enterobacter cloacae bloodstream infections and PR at Hôpital Charles-LeMoyne, a 590-bed university-affiliated teaching hospital in the Montreal area, that was first detected in August 1995 when five episodes of bacteremia occurred among the dialysis unit's 60 chronic hemodialysis patients.…”
Section: Introductionmentioning
confidence: 99%
“…Periodical and long activation of blood cells by HD is now recognized as the major cause of chronic microinflammation that is implied in the development of long-term dialysis-related complications. So it has been generally recognized that contaminated dialysis fluid was one of the most important factors which induced dialysis-related complications such as dialysisrelated amyloidosis (DRA) [4] , anemia [5] , cardiovascular complications [6] , intradialytic complications [7] , neurologic disorders [8] , nutrition [9] , residual renal function [10] and poor prognosis of dialysis patients ( table 1 ). From the point of view for preventing inflammation in most dialysis-related pathologies, the standard recommends that UPD should be used for all dialysis modalities at all dialysis facilities.…”
Section: Current Situation Of Endotoxin Retentive Filtermentioning
confidence: 99%
“…cultured using calibrated loop and blood agar plates-results were always recorded as no growth Phialemonium curvatum access infections Observations at the facility noted some irregularities Review infection control practices; clean and disinfect in 4 hemodialysis patients; 2 of these in site prep for needle insertion. All 156 of headers with gauze; staff not routinely header; change gloves frequently; process dialyzers changing gloves; dialyzers not processed for immediately after rinsing and cleaning several hours after disassembly and cleaning Pyrogenic reactions in 3 high-flux dialysis patients 157 Dialyzer reprocessed with two disinfectants; water Do not disinfect dialyzers with multiple germicides; more used for reuse did not meet AAMI standards frequent disinfection of water system Pyrogenic reactions in 14 high-flux dialysis Dialyzers rinsed with city water; water for reuse Do not rinse or clean dialyzers with city water; disinfect patients; 1 death 158 did not meet AAMI standards water treatment system more frequently Pyrogenic reactions in 18 patients 51 Dialyzers rinsed with city water containing high Do not rinse or clean dialyzers with city water; disinfect levels of endotoxin; water used for reuse did not water treatment system more frequently meet AAMI standards Pyrogenic reactions in 22 patients 159 Water for reuse did not meet AAMI standards; Use correct microbial assay procedure; disinfect water improper microbial assay technique distribution system…”
Section: High-flux Dialysis and Bicarbonate Dialysatementioning
confidence: 97%