2006
DOI: 10.1016/j.nima.2006.06.024
|View full text |Cite
|
Sign up to set email alerts
|

A pixel ionization chamber used as beam monitor at the Institut Curie—Centre de Protontherapie de Orsay (CPO)

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

0
13
0

Year Published

2007
2007
2017
2017

Publication Types

Select...
7
1

Relationship

5
3

Authors

Journals

citations
Cited by 30 publications
(13 citation statements)
references
References 6 publications
0
13
0
Order By: Relevance
“…The size of these detectors is 24 × 24 cm 2 , to cover the maximum beam size. The detector front-end readout is based on custom designed boards, which host application specific integrated circuit (ASIC) electronics custom designed for this purpose [15,16]. The readout frequency for the integral chamber is about 1 MHz, while for the strip chambers 20 and 10 kHz respectively for x and y strips.…”
Section: Detectors For Beam Control 21 Ionization Chambersmentioning
confidence: 99%
“…The size of these detectors is 24 × 24 cm 2 , to cover the maximum beam size. The detector front-end readout is based on custom designed boards, which host application specific integrated circuit (ASIC) electronics custom designed for this purpose [15,16]. The readout frequency for the integral chamber is about 1 MHz, while for the strip chambers 20 and 10 kHz respectively for x and y strips.…”
Section: Detectors For Beam Control 21 Ionization Chambersmentioning
confidence: 99%
“…Their design is similar to other detectors successfully developed by the Torino group for different medical applications [5,6]. For safety and redundancy requirements, two independent detectors have been designed for each treatment line: they are housed in separate boxes (called BOX1 and BOX2) and are read out and processed independently during the treatment.…”
Section: Detector Descriptionmentioning
confidence: 99%
“…As described in [5], the electrodes are stretched and glued on fiberglass frames which also provide mechanical support for the front-end readout boards. Additional frames are added between the electrodes to provide a gas gap of 5 mm for each chamber.…”
Section: Detector Descriptionmentioning
confidence: 99%
“…Most of the beam monitor detectors are based on ionization chambers [6,7] that offer several advantages in terms of simplicity, robustness and minimal perturbation of the beam. In the perspective of these emerging technologies, the chamber efficiency has to be monitored and eventually corrected for.…”
Section: Introductionmentioning
confidence: 99%
“…In this paper we present a simple and effective solution to increase by almost two orders of magnitude the input current range of an Application Specific Integrated Circuit (ASIC) which was designed by our group and used in several laboratories worldwide [6][7][8], still maintaining a good performance in terms of noise and resolution. The chip, named TERA, implements an architecture based on a charge-to-count converter and integrates 64 equal channels.…”
Section: Introductionmentioning
confidence: 99%