We developed a significantly improved ultra-high performance liquid chromatography-tandem mass spectrometry method for determination of 5-nitro-2-furaldehyde (NF) as a surrogate using a novel internal standard for the detection of nitrofurazone. We used 2,4-dinitrophenylhydrazine derivatization and furfural as the internal standard. Derivatization was easily performed in HCl using ultrasonic manipulation for 5 min followed by liquid extraction using ethyl acetate. The samples were concentrated and purified using reverse phase and alumina cartridges in tandem. The derivatives were separated using a linear gradient elution on a C18 column with methanol and water as the mobile phase in negative ionization mode and multiple reaction monitoring. Under the optimized conditions, the calibration curves were linear from 0.2 to 20 μg/L with correlation coefficients >0.999. Mean recoveries were 80.8 to 104.4% with the intra- and inter-day relative standard deviations <15% at spiking levels of 0.1 to 10 μg/kg. The limits of detection and quantification were 0.05 and 0.1 μg/kg, respectively. This method is a robust tool for the identification and quantitative determination of NF in shrimp samples.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.