Abstract:An optimized method based on GC-MS with ethyl chloroformate derivatization has been developed for the comprehensive analysis of endogenous metabolites in serum. Twenty-two reference standards and serum samples were used to validate the proposed method. The correlation coefficient was higher than 0.9900 for each of the standards, and the LOD varied from 125 to 300 pg on-column. The analytical equipment exhibited good repeatability (RSD<10%) for all of the standards. Both the repeatability and the within-48-h stability of the analytical method were satisfactory (RSD<10%) for the 18 metabolites identified in the serum samples. Mean recovery was acceptable for the 18 metabolites, ranging from 70% to 120% with RSDs of less than 10%. Using the optimized protocol and a subsequent multivariate statistical technique, complete differentiation was achieved between the metabolic profile of uremic patients and that of age-and sex-matched normal subjects. Significantly decreased levels of valine, leucine, and isoleucine and increased levels of myristic acid and linoleic acid were observed in the patient group. This work demonstrated that this method is suitable for serumbased metabolic profiling studies.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.